A low-kilovolt (peak) high-tube current technique improves venous enhancement and reduces the radiation dose at indirect multidetector-row CT venography: initial experience.
Nakaura, Takeshi; Awai, Kazuo; Oda, Seitaro; Yanaga, Yumi; Namimoto, Tomohiro; Harada, Kazunori; Uemura, Shouzaburou; Yamashita, Yasuyuki.
Afiliação
  • Nakaura T; Diagnostic Radiology, Amakusa Medical Center, Amakusa, Japan. kff00712@nifty.com
J Comput Assist Tomogr ; 35(1): 141-7, 2011.
Article em En | MEDLINE | ID: mdl-21245699
ABSTRACT

OBJECTIVE:

The purpose was to evaluate the image noise, contrast-to-noise ratio, and radiation dose at indirect multidetector-row computed tomographic venography (CTV) using a low-kilovolt (peak) (kV[p]) high-tube current technique. MATERIALS AND

METHODS:

Thirty patients underwent indirect CTV of the lower extremity at 120 kV(p) with 170 mA and at 80 kV(p) with 426 mA (CT dose index volumes 11.8 and 9.0 mGy, respectively) on a 64-detector CT scanner. We assessed CT numbers, image noise, and contrast-to-noise ratio of deep (femoral and popliteal) veins.

RESULTS:

The mean (SD) CT numbers of the femoral and popliteal veins were significantly greater at 80 kV(p) than at 120 kV(p) (142.8 [24.5] and 147.0 [19.4] Hounsfield units [HU], respectively, vs 93.1 [15.5] and 105.0 [14.9] HU, respectively; P < 0.01). The mean (SD) image noise was significantly higher at 80 kV(p) than at 120 kV(p) (11.6 [2.6] and 7.4 [1.4] HU, respectively, vs 8.9 [1.6] and 6.2 [1.5] HU, respectively; P < 0.01). Contrast-to-noise ratios at the femoral and popliteal veins were significantly higher at 80 kV(p) than at 120 kV(p) (6.0 [2.2] and 9.7 [2.4], respectively vs 3.3 [1.8] and 6.9 [2.5], respectively; P < 0.01).

CONCLUSION:

At indirect CTV, 80 kV(p) and a high tube current setting yielded significantly improved image quality at a reduced radiation dose compared with 120 kV(p).
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Veia Poplítea / Doses de Radiação / Angiografia / Tomografia Computadorizada por Raios X / Extremidade Inferior / Veia Femoral Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2011 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Veia Poplítea / Doses de Radiação / Angiografia / Tomografia Computadorizada por Raios X / Extremidade Inferior / Veia Femoral Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2011 Tipo de documento: Article