Helmholtz-pair transmit coil with integrated receive array for high-resolution MRI of trabecular bone in the distal tibia at 7T.
Wright, Alexander C; Lemdiasov, Rostislav; Connick, Thomas J; Bhagat, Yusuf A; Magland, Jeremy F; Song, Hee Kwon; Toddes, Steven P; Ludwig, Reinhold; Wehrli, Felix W.
Afiliação
  • Wright AC; Department of Radiology, University of Pennsylvania Medical Center, Philadelphia, PA, USA. Alexander.Wright@uphs.upenn.edu
J Magn Reson ; 210(1): 113-22, 2011 May.
Article em En | MEDLINE | ID: mdl-21402488
A Helmholtz-pair local transmit RF coil with an integrated four-element receive array RF coil and foot immobilization platform was designed and constructed for imaging the distal tibia in a whole-body 7T MRI scanner. Simulations and measurements of the B(1) field distribution of the transmit coil are described, along with SAR considerations for operation at 7T. Results of imaging the trabecular bone of three volunteers at 1.5T, 3T and 7T are presented, using identical 1.5T and 3T versions of the 7T four-element receive array. The spatially registered images reveal improved visibility for individual trabeculae and show average gains in SNR of 2.8× and 4.9× for imaging at 7T compared to 3T and 1.5T, respectively. The results thus display an approximately linear dependence of SNR with field strength and enable the practical utility of 7T scanners for micro-MRI of trabecular bone.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tíbia / Imageamento por Ressonância Magnética Tipo de estudo: Diagnostic_studies Limite: Adult / Humans / Male Idioma: En Ano de publicação: 2011 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tíbia / Imageamento por Ressonância Magnética Tipo de estudo: Diagnostic_studies Limite: Adult / Humans / Male Idioma: En Ano de publicação: 2011 Tipo de documento: Article