Your browser doesn't support javascript.
loading
Reproducibility and robustness of a real-time microfluidic cell toxicity assay.
Cooksey, Gregory A; Elliott, John T; Plant, Anne L.
Afiliação
  • Cooksey GA; Biochemical Science Division, NIST, Gaithersburg, Maryland, USA. gregory.cooksey@nist.gov
Anal Chem ; 83(10): 3890-6, 2011 May 15.
Article em En | MEDLINE | ID: mdl-21506521
ABSTRACT
Numerous opportunities exist to apply microfluidic technology to high-throughput and high-content cell-based assays. However, maximizing the value of microfluidic assays for applications such as drug discovery, screening, or toxicity evaluation will require assurance of within-device repeatability, day-to-day reproducibility, and robustness to variations in conditions that might occur from laboratory to laboratory. This report describes a study of the performance and variability of a cell-based toxicity assay in microfluidic devices made of poly(dimethylsiloxane) (PDMS). The assay involves expression of destabilized green fluorescent protein (GFP) as a reporter of intracellular protein synthesis and degradation. Reduction in cellular GFP due to inhibition of ribosome activity by cycloheximide (CHX) was quantified with real-time quantitative fluorescence imaging. Assay repeatability was measured within a 64-chamber microfluidic device. Assay performance across a range of cell loading densities within a single device was assessed, as was replication of measurements in microfluidic devices prepared on different days. Assay robustness was tested using different fluorescence illumination sources and reservoir-to-device tubing choices. Both microfluidic and larger scale assay conditions showed comparable GFP decay rates upon CHX exposure, but the microfluidic data provided the higher level of confidence.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Testes de Toxicidade / Técnicas Analíticas Microfluídicas Limite: Animals Idioma: En Ano de publicação: 2011 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Testes de Toxicidade / Técnicas Analíticas Microfluídicas Limite: Animals Idioma: En Ano de publicação: 2011 Tipo de documento: Article