Your browser doesn't support javascript.
loading
A beam-specific planning target volume (PTV) design for proton therapy to account for setup and range uncertainties.
Park, Peter C; Zhu, X Ronald; Lee, Andrew K; Sahoo, Narayan; Melancon, Adam D; Zhang, Lifei; Dong, Lei.
Afiliação
  • Park PC; Medical Physics Program, Graduate School of Biomedical Sciences, The University of Texas Health Science Center at Houston, TX, USA.
Int J Radiat Oncol Biol Phys ; 82(2): e329-36, 2012 Feb 01.
Article em En | MEDLINE | ID: mdl-21703781
ABSTRACT

PURPOSE:

To report a method for explicitly designing a planning target volume (PTV) for treatment planning and evaluation in heterogeneous media for passively scattered proton therapy and scanning beam proton therapy using single-field optimization (SFO). METHODS AND MATERIALS A beam-specific PTV (bsPTV) for proton beams was derived by ray-tracing and shifting ray lines to account for tissue misalignment in the presence of setup error or organ motion. Range uncertainties resulting from inaccuracies in computed tomography-based range estimation were calculated for proximal and distal surfaces of the target in the beam direction. The bsPTV was then constructed based on local heterogeneity. The bsPTV thus can be used directly as a planning target as if it were in photon therapy. To test the robustness of the bsPTV, we generated a single-field proton plan in a virtual phantom. Intentional setup and range errors were introduced. Dose coverage to the clinical target volume (CTV) under various simulation conditions was compared between plans designed based on the bsPTV and a conventional PTV.

RESULTS:

The simulated treatment using the bsPTV design performed significantly better than the plan using the conventional PTV in maintaining dose coverage to the CTV. With conventional PTV plans, the minimum coverage to the CTV dropped from 99% to 67% in the presence of setup error, internal motion, and range uncertainty. However, plans using the bsPTV showed minimal drop of target coverage from 99% to 94%.

CONCLUSIONS:

The conventional geometry-based PTV concept used in photon therapy does not work well for proton therapy. We investigated and validated a beam-specific PTV method for designing and evaluating proton plans.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Planejamento da Radioterapia Assistida por Computador / Incerteza / Carga Tumoral / Erros de Configuração em Radioterapia / Terapia com Prótons / Movimento Limite: Humans Idioma: En Ano de publicação: 2012 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Planejamento da Radioterapia Assistida por Computador / Incerteza / Carga Tumoral / Erros de Configuração em Radioterapia / Terapia com Prótons / Movimento Limite: Humans Idioma: En Ano de publicação: 2012 Tipo de documento: Article