Your browser doesn't support javascript.
loading
HPLC-MS/MS method for the determination of cytarabine in human plasma.
Hilhorst, Martijn J; Hendriks, Gert; van Hout, Mischa W J; Sillén, Henrik; van de Merbel, Nico C.
Afiliação
  • Hilhorst MJ; PRA International Early Development Services, Assen, The Netherlands. hilhorstmartijn@praintl.com
Bioanalysis ; 3(14): 1603-11, 2011 Jul.
Article em En | MEDLINE | ID: mdl-21756093
ABSTRACT

BACKGROUND:

Cytarabine is an anti-tumor drug that is currently under investigation for treatment in combination with other anticancer drugs for the chemotherapy of leukemia. The quantitative determination of cytarabine in plasma is challenging due to the required sensitivity, its in vitro instability and the presence of an isobaric endogenous compound, cytidine. Owing to the polarity of cytarabine, conventional chromatography cannot provide adequate separation of the analyte and the interfering compounds. A few analytical methods have been reported for the determination of cytarabine in plasma, but their sensitivity was not sufficient since most of these methods apply spectrophotometric detection.

RESULTS:

In this article, an LC-MS/MS method is described for the determination of cytarabine in human plasma down to the sub ng/ml level. To prevent conversion of cytarabine by cytidine deaminase, whole blood samples were stabilized with tetrahydrouridine directly after the collection of whole blood at the clinical site. Cation-exchange SPE was employed to extract cytarabine from 50 µl human plasma. UHPLC on high strength silica T3 column (100 × 2.1 mm × 1.8 µm) was applied to achieve adequate separation of cytarabine from cytidine within a reasonable run time of 5 min. A triple quad mass spectrometer equipped with an ESI source was used for detection.

CONCLUSION:

The method was linear over the concentration ranges of 0.500-500 ng/ml. The intra- and inter-day relative standard deviation (precision) as well as the bias (accuracy) were well below 15%. In the presence of tetrahydrouridine, cytarabine was sufficiently stable under all relevant conditions. The method was successfully applied to support a clinical pharmacokinetic study with a low dose of cytarabine.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Cromatografia Líquida de Alta Pressão / Cromatografia Líquida / Citarabina / Espectrometria de Massas em Tandem / Antimetabólitos Antineoplásicos Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Ano de publicação: 2011 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Cromatografia Líquida de Alta Pressão / Cromatografia Líquida / Citarabina / Espectrometria de Massas em Tandem / Antimetabólitos Antineoplásicos Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Ano de publicação: 2011 Tipo de documento: Article