Your browser doesn't support javascript.
loading
First order correction for T2*-relaxation in determining contrast agent concentration from spoiled gradient echo pulse sequence signal intensity.
De Naeyer, Dieter; Debergh, Isabelle; De Deene, Yves; Ceelen, Wim P; Segers, Patrick; Verdonck, Pascal.
Afiliação
  • De Naeyer D; Institute Biomedical Technology (IBiTech), Department of Civil Engineering, Ghent University, Belgium. dieter.denaeyer@ugent.be
J Magn Reson Imaging ; 34(3): 710-5, 2011 Sep.
Article em En | MEDLINE | ID: mdl-21769976
ABSTRACT

PURPOSE:

To investigate the accuracy of a method neglecting T(2)*-relaxation, for the conversion of spoiled gradient echo pulse sequence signal intensity to contrast agent (CA) concentration, in dynamic contrast enhanced MRI studies. In addition a new closed form conversion expression is proposed that accounts for a first order approximation of T(2)*-relaxation. MATERIALS AND

METHODS:

The accuracy of both conversion methods is compared theoretically by means of simulations for four pulse sequences from literature. Both methods are tested in vivo against the numerical conversion method for measuring the arterial input function in mice.

RESULTS:

Simulations show that the T(2)*-neglecting method underestimates typical tissue CA concentrations (0 mM to 2 mM) up to 6%, while the errors for arterial concentrations (0 mM to 10 mM) range up to 43%. The results from our first order method are numerically indistinguishable from the simulation input values in tumor tissue, while for arterial concentrations the error is reduced up to a factor 10. In vivo, peak Gd-DOTA concentration is underestimated up to 14% with the T(2)*-neglecting method and up to 0.9% with our first order method.

CONCLUSION:

Our conversion method reduces the underestimation of CA concentration severely in a broad physiological concentration range and is easy to perform in any clinical setting.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Compostos Organometálicos / Algoritmos / Interpretação de Imagem Assistida por Computador / Artefatos / Imagem Ecoplanar / Compostos Heterocíclicos Tipo de estudo: Diagnostic_studies Limite: Animals Idioma: En Ano de publicação: 2011 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Compostos Organometálicos / Algoritmos / Interpretação de Imagem Assistida por Computador / Artefatos / Imagem Ecoplanar / Compostos Heterocíclicos Tipo de estudo: Diagnostic_studies Limite: Animals Idioma: En Ano de publicação: 2011 Tipo de documento: Article