Asymmetric logistic peak as a suitable function for the resolution of highly asymmetric voltammograms in non-bilinear systems.
Analyst
; 136(22): 4696-703, 2011 Nov 21.
Article
em En
| MEDLINE
| ID: mdl-21961109
The asymmetric logistic peak is tested as a new function for the parametric signal fitting (PSF) of highly asymmetric electrochemical signals in non-bilinear datasets, such as those obtained in linear sweep voltammetry (LSV) or in the presence of irreversible electrochemical processes. This new multivariate curve resolution strategy (PSF-ALPA) is successfully applied to LS voltammograms measured for the Cd(II)-glutathione system with a hanging mercury drop electrode, where Cd(II) is reversibly reduced, and to differential pulse voltammograms (DPV) measured at a glassy carbon electrode, where Cd(II) reduction becomes irreversible. Matrix augmentation by using LS voltammograms measured at different scan rates provides good results and encourages the development of ALPA methodology for third order data.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Idioma:
En
Ano de publicação:
2011
Tipo de documento:
Article