Calculating the peak skin dose resulting from fluoroscopically-guided interventions. Part II: case studies.
Jones, A Kyle; Pasciak, Alexander S.
Afiliação
  • Jones AK; Department of Imaging Physics, Division of Diagnostic Imaging, The University of Texas MD Anderson Cancer Center, Houston, TX 77030, USA. kyle.jones@mdanderson.org
J Appl Clin Med Phys ; 13(1): 3693, 2012 Jan 05.
Article em En | MEDLINE | ID: mdl-22231220
ABSTRACT
Part II of this review article presents four case studies for which peak skin doses are calculated following the methods outlined in Part I of this review. The data available in the cases ranges from proprietary dose reports to fluoroscopy time and number of digital acquisition frames only. Flowcharts are provided for each case. These flowcharts outline the calculation steps and data sources used to estimate the peak skin dose. The accuracy that can be achieved using these methods depends on several factors, including the calibration accuracy of dosimetric equipment, accuracy of information reported in the DICOM header and proprietary dose reports, accuracy of quantities measured by the medical physicist, and procedural factors such as rotation of the C-arm during a fluoroscopically-guided procedure.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Radiometria / Fenômenos Fisiológicos da Pele / Fluoroscopia / Cirurgia Assistida por Computador / Modelos Biológicos Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Ano de publicação: 2012 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Radiometria / Fenômenos Fisiológicos da Pele / Fluoroscopia / Cirurgia Assistida por Computador / Modelos Biológicos Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Ano de publicação: 2012 Tipo de documento: Article