Prevention of inappropriate ICD shocks due to lead insulation failure by continuous monitoring and automatic alert.
Gelder, Robert N; Gunderson, Bruce D.
Afiliação
  • Gelder RN; Department of Medicine, Victorian Heart Centre, Epworth Hospital, Melbourne, Australia. rgelder@yahoo.com
Pacing Clin Electrophysiol ; 35(6): e150-3, 2012 Jun.
Article em En | MEDLINE | ID: mdl-22309317
ABSTRACT
Patients with implantable cardioverter defibrillator lead insulation failures may present with oversensing and/or abnormal impedance. The Lead Integrity Alert (LIA) monitors right ventricular pace/sense leads using both continuous oversensing and daily impedance measurementd. Oversensing consists of isolated short R-R intervals and nonsustained runs of short R-R intervals. The LIA algorithm has been studied for Sprint Fidelis conductor fractures, but not for lead insulation failures. We report on a patient with a failed St. Jude Riata™ ST lead (St. Jude Medical, St. Paul, MN, USA) connected to a Medtronic Virtuoso DR (Medtronic Inc., Minneapolis, MN, USA) with the LIA. Oversensing triggered the LIA, while the impedance trend was normal.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Eletrocardiografia Ambulatorial / Desfibriladores Implantáveis / Traumatismos por Eletricidade / Eletrodos Implantados / Falha de Equipamento / Alarmes Clínicos Limite: Adult / Humans / Male Idioma: En Ano de publicação: 2012 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Eletrocardiografia Ambulatorial / Desfibriladores Implantáveis / Traumatismos por Eletricidade / Eletrodos Implantados / Falha de Equipamento / Alarmes Clínicos Limite: Adult / Humans / Male Idioma: En Ano de publicação: 2012 Tipo de documento: Article