Optical coherence tomography for whole eye segment imaging.
Dai, Cuixia; Zhou, Chuanqing; Fan, Shanhui; Chen, Zhe; Chai, Xinyu; Ren, Qiushi; Jiao, Shuliang.
Afiliação
  • Dai C; School of Biomedical Engineering, Shanghai Jiao Tong University, 800 Dongchuan Road, Shanghai 200240, China.
Opt Express ; 20(6): 6109-15, 2012 Mar 12.
Article em En | MEDLINE | ID: mdl-22418490
ABSTRACT
We proposed a dual focus dual channel spectral domain optical coherence tomography (SD-OCT) for simultaneous imaging of the whole eye segments from cornea to the retina. By using dual channels the system solved the problem of limited imaging depth of SD-OCT. By using dual focus the system solved the problem of simultaneous light focusing on the anterior segment of the eye and the retina. Dual focusing was achieved by adjusting the collimating lenses so the divergence of the two probing beams was tuned to make them focused at different depth in the eye. We further achieved full range complex (FRC) SD-OCT in one channel to increase the depth range for anterior segment imaging. The system was successfully tested by imaging a human eye in vivo.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Aumento da Imagem / Oftalmoscópios / Tomografia de Coerência Óptica / Olho Tipo de estudo: Diagnostic_studies / Evaluation_studies Limite: Humans Idioma: En Ano de publicação: 2012 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Aumento da Imagem / Oftalmoscópios / Tomografia de Coerência Óptica / Olho Tipo de estudo: Diagnostic_studies / Evaluation_studies Limite: Humans Idioma: En Ano de publicação: 2012 Tipo de documento: Article