Evaluation of the rapid diagnostic test CareStart pLDH Malaria (Pf-pLDH/pan-pLDH) for the diagnosis of malaria in a reference setting.
Heutmekers, Marloes; Gillet, Philippe; Maltha, Jessica; Scheirlinck, Annelies; Cnops, Lieselotte; Bottieau, Emmanuel; Van Esbroeck, Marjan; Jacobs, Jan.
Afiliação
  • Heutmekers M; Maastricht University, Maastricht, The Netherlands.
Malar J ; 11: 204, 2012 Jun 18.
Article em En | MEDLINE | ID: mdl-22704733
ABSTRACT

BACKGROUND:

The present study evaluated CareStart pLDH Malaria, a three-band rapid diagnostic test detecting Plasmodium falciparum-specific parasite lactate dehydrogenase (Pf-pLDH) and pan Plasmodium-specific pLDH (pan-pLDH) in a reference setting.

METHODS:

CareStart pLDH was retrospectively and prospectively assessed with a panel of stored (n=498) and fresh (n=77) blood samples obtained in international travelers suspected of malaria. Both panels comprised all four Plasmodium species; the retrospective panel comprised also Plasmodium negative samples. The reference method was microscopy corrected by PCR. The prospective panel was run side-to-side with OptiMAL (Pf-pLDH/pan-pLDH) and SDFK60 (histidine-rich protein-2 (HRP-2)/pan-pLDH).

RESULTS:

In the retrospective evaluation, overall sensitivity for P. falciparum samples (n=247) was 94.7%, reaching 98.7% for parasite densities>1,000/µl. Most false negative results occurred among samples with pure gametocytaemia (2/12, 16.7%) and at parasite densities ≤ 100/µl (7/12, 58.3%). None of the Plasmodium negative samples (n=96) showed visible test lines. Sensitivities for Plasmodium vivax (n=70), Plasmodium ovale (n=69) and Plasmodium malariae (n=16) were 74.3%, 31.9% and 25.0% respectively. Wrong species identification occurred in 10 (2.5%) samples and was mainly due to P. vivax samples reacting with the Pf-pLDH test line. Overall, Pf-pLDH test lines showed higher line intensities compared to the pan-pLDH lines (67.9% and 23.0% medium and strong line intensities for P. falciparum). In the prospective panel (77 Plasmodium-positive samples), CareStart pLDH showed higher sensitivities for P. falciparum compared to OptiMAL (p=0.008), lower sensitivities for P. falciparum as compare to SDFK60 (although not reaching statistical significance, p=0.08) and higher sensitivities for P. ovale compared to both OptiMAL (p=0.03) and SDFK60 (p=0.01). Inter-observer and test reproducibility were good to excellent.

CONCLUSION:

CareStart pLDH performed excellent for the detection of P. falciparum, well for P. vivax, but poor for P. ovale and P. malariae.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Plasmodium / Testes Diagnósticos de Rotina / Malária Tipo de estudo: Diagnostic_studies / Evaluation_studies / Prognostic_studies Limite: Adolescent / Adult / Aged / Child / Child, preschool / Female / Humans / Infant / Male / Middle aged Idioma: En Ano de publicação: 2012 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Plasmodium / Testes Diagnósticos de Rotina / Malária Tipo de estudo: Diagnostic_studies / Evaluation_studies / Prognostic_studies Limite: Adolescent / Adult / Aged / Child / Child, preschool / Female / Humans / Infant / Male / Middle aged Idioma: En Ano de publicação: 2012 Tipo de documento: Article