Your browser doesn't support javascript.
loading
A comparison of hexaminolevulinate (Hexvix(®)) fluorescence cystoscopy and white-light cystoscopy for detection of bladder cancer: results of the HeRo observational study.
Lapini, Alberto; Minervini, Andrea; Masala, Alberto; Schips, Luigi; Pycha, Armin; Cindolo, Luca; Giannella, Riccardo; Martini, Thomas; Vittori, Gianni; Zani, Danilo; Bellomo, Fania; Cosciani Cunico, Sergio.
Afiliação
  • Lapini A; Department of Urology, Careggi Hospital, University of Florence, Florence, Italy.
Surg Endosc ; 26(12): 3634-41, 2012 Dec.
Article em En | MEDLINE | ID: mdl-22729704
ABSTRACT

BACKGROUND:

To date, no study has presented results of photodynamic diagnosis (PDD) cystoscopy compared with white-light cystoscopy (WLC) in daily practice. The aim of the present study is to evaluate the diagnostic accuracy of hexylaminolevulinate hydrochloride (Hexvix(®)) PDD cystoscopy compared with standard WLC used in daily practice.

METHODS:

An observational, open-label, comparative, controlled (within patient), multicenter study was carried out on 96 consecutive patients with suspected or confirmed bladder cancer. All patients had standard WLC followed by blue-light cystoscopy (BLC). Positive lesions detected using WLC and BLC were recorded. Biopsies/resection of each positive lesion were taken after the bladder was inspected. Sensitivity, specificity, positive predictive value, and negative predictive value with each method were calculated.

RESULTS:

Overall, 234 suspicious lesions were detected; 108 (46.2%) were histologically confirmed to be bladder tumors/carcinoma in situ (CIS). The sensitivity of BLC biopsies was significantly higher than for WLC technique (99.1 vs 76.8%; p < 0.00001). The relative sensitivity of BLC versus WLC was 1.289, showing superiority of BLC of 28.9%. The specificity of BLC biopsies was not significantly different compared with WLC (36.5 vs 30.2%). Positive predictive value for BLC- and WLC-guided biopsies was 54.9 and 50.9%, respectively. Negative predictive value per biopsy for BLC- and WLC-guided biopsies was 97.4 and 64.8%, respectively. BLC and WLC reached the correct diagnosis in 97.9 and 88.5% of patients, respectively. This difference was statistically significant (p = 0.0265). The lack of a random biopsy protocol was the major limitation of the study.

CONCLUSIONS:

Hexvix(®) PDD cystoscopy used in daily practice enhances the diagnostic accuracy of standard cystoscopy with higher negative predictive value, potentially permitting an improvement in patient prognosis.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Neoplasias da Bexiga Urinária / Cistoscopia / Ácido Aminolevulínico / Luz Tipo de estudo: Clinical_trials / Diagnostic_studies / Guideline / Observational_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2012 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Neoplasias da Bexiga Urinária / Cistoscopia / Ácido Aminolevulínico / Luz Tipo de estudo: Clinical_trials / Diagnostic_studies / Guideline / Observational_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2012 Tipo de documento: Article