Your browser doesn't support javascript.
loading
High-resolution Thomson scattering system on the COMPASS tokamak: evaluation of plasma parameters and error analysis.
Aftanas, M; Bohm, P; Bilkova, P; Weinzettl, V; Zajac, J; Zacek, F; Stockel, J; Hron, M; Panek, R; Scannell, R; Walsh, M J.
Afiliação
  • Aftanas M; Institute of Plasma Physics AS CR, vvi, Prague 8, Czech Republic. aftanas@ipp.cas.cz
Rev Sci Instrum ; 83(10): 10E350, 2012 Oct.
Article em En | MEDLINE | ID: mdl-23127007
ABSTRACT
The electron density and temperature profiles measured by the Thomson scattering diagnostic on the COMPASS tokamak are used for estimation of electron kinetic energy, energy confinement time, and effective charge number Z(eff). Data are compared with the line-integrated electron density measured by a microwave interferometer in an ohmically heated plasma with a circular cross section. An error analysis of both electron temperature and density are performed by two methods-a constant chi-square boundaries method and a Monte Carlo simulation, determining asymmetrical error bars for the electron temperature.

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Ano de publicação: 2012 Tipo de documento: Article