The performances of standard and ResMed masks during bag-valve-mask ventilation.
Lee, Hyoung Youn; Jeung, Kyung Woon; Lee, Byung Kook; Lee, Seung Joon; Jung, Yong Hun; Lee, Geo Sung; Min, Yong Il; Heo, Tag.
Afiliação
  • Lee HY; Department of Emergency Medicine, Chonnam National University Hospital, Gwangju, Republic of Korea.
Prehosp Emerg Care ; 17(2): 235-40, 2013.
Article em En | MEDLINE | ID: mdl-23157368
BACKGROUND: A tight mask seal is frequently difficult to obtain and maintain during single-rescuer bag-valve-mask (BVM) ventilation. The ResMed mask (Bella Vista, NSW, Australia) is a continuous-positive-airway-pressure mask (CM) designed for noninvasive ventilation. OBJECTIVE: In this study, we compared the ventilation performances of a standard mask (SM) and a ResMed CM using a simulation manikin in an out-of-hospital single-rescuer BVM ventilation scenario. METHODS: Thirty emergency medical technicians (EMTs) performed two 2-minute attempts to ventilate a simulation manikin using BVM ventilation, alternatively, with the SM or the ResMed CM in a randomized order. Ventilation parameters including tidal volume and peak airway pressure were measured using computer analysis software connected to the simulation manikin. Successful volume delivery was defined as delivery of 440-540 mL of tidal volume in accord with present cardiopulmonary resuscitation guidelines. RESULTS: BVM ventilation using the ResMed CM produced higher mean (± standard deviation) tidal volumes (452 ± 50 mL vs. 394 ± 113 mL, p = 0.014) and had a higher proportion of successful volume deliveries (65.3% vs. 26.7%, p < 0.001) than that using the SM. Peak airway pressure was higher in BVM ventilation using the ResMed CM (p = 0.035). Stomach insufflation did not occur during either method. Twenty-nine of the participants (96.7%) preferred BVM ventilation using the ResMed CM. CONCLUSIONS: BVM ventilations using ResMed CM resulted in a significantly higher proportion of successful volume deliveries meeting the currently recommended range of tidal volume. Clinical studies are needed to determine the value of the ResMed CM for BVM ventilation.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Máscaras Laríngeas / Pressão Positiva Contínua nas Vias Aéreas Tipo de estudo: Clinical_trials / Guideline / Observational_studies Limite: Adult / Female / Humans / Male País/Região como assunto: Asia Idioma: En Ano de publicação: 2013 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Máscaras Laríngeas / Pressão Positiva Contínua nas Vias Aéreas Tipo de estudo: Clinical_trials / Guideline / Observational_studies Limite: Adult / Female / Humans / Male País/Região como assunto: Asia Idioma: En Ano de publicação: 2013 Tipo de documento: Article