Difference in dose area product between analog image intensifier and digital flat panel detector in peripheral angiography and the effect of BMI.
Wiesinger, B; Kirchner, S; Blumenstock, G; Herz, K; Schmehl, J; Claussen, C D; Wiskirchen, J.
Afiliação
  • Wiesinger B; Diagnostische und Interventionelle Radiologie, Eberhard-Karls-Universitätsklinikum Tübingen. benjamin.wiesinger@med.unituebingen.de
Rofo ; 185(2): 153-9, 2013 Feb.
Article em En | MEDLINE | ID: mdl-23196837
ABSTRACT

PURPOSE:

Comparison of dose area products (DAP) in diagnostic angiography procedures between an image intensifier (II) and a flat panel detector (FPD) angiography system and the evaluation of DAP/body mass index (BMI) dependency. MATERIALS AND

METHODS:

An image intensifier system or a flat panel detector system was used to perform 571 diagnostic angiographies (n = 328 and n = 243, respectively) of 5 different types peripheral arterial, venous, single leg, abdominal and upper extremity. The results were retrospectively analyzed. The DAP, fluoroscopy time (t) and the number of series of the respective interventions as calculated by the respective machines was compared for all interventions and for the respective subtypes and machines. The BMI dependency was calculated separately for both machines for all interventions by subdividing the patients into 6 BMI classes defined by the WHO.

RESULTS:

The average DAP for all diagnostic interventions was 1958.9 cGy×cm2 (t = 384.6 s, n = 7.85 series) for the II and 2927.4 cGy×cm2 (t = 267.4 s, n = 7.02 series) for the FPD. Group-dependent differences ranged between + 21 and + 252 % when using the FPD system. After time standardization, the respective increases were found to be 120 % for the FPD system. The DAPs increased considerably in patients with higher BMIs (766.7 cGy × cm2 - 6892.6 cGy × cm2, II machine, 950.5 cGy × cm2 - 12 487.7 cGy × cm2, FPD machine) with a greater DAP gain seen for the FPD. The average duration of the interventions was higher using the II machine.

CONCLUSION:

The use of an FPD system led to higher DAP values compared to the II system in diagnostic angiographic procedures. In addition, increased BMI values led to higher DAPs, especially for the FPD machine. However, the average fluoroscopy times were shorter.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Proteção Radiológica / Ecrans Intensificadores para Raios X / Angiografia / Intensificação de Imagem Radiográfica / Índice de Massa Corporal / Doença Arterial Periférica Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Ano de publicação: 2013 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Proteção Radiológica / Ecrans Intensificadores para Raios X / Angiografia / Intensificação de Imagem Radiográfica / Índice de Massa Corporal / Doença Arterial Periférica Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Ano de publicação: 2013 Tipo de documento: Article