Fully automated reagent peak-free liquid chromatography fluorescence analysis of highly polar carboxylic acids using a column-switching system and fluorous scavenging derivatization.
J Sep Sci
; 36(2): 232-8, 2013 Jan.
Article
em En
| MEDLINE
| ID: mdl-23229764
In this study, we combined a column-switching system with a fluorous scavenging derivatization method to develop a fully automated reagent peak-free LC fluorescence detection protocol for the analysis of highly polar carboxylic acids. In this method, highly polar carboxylic acids were derivatized with fluorescent 1-pyrenemethylamine in the presence of 1-ethyl-3-(3-dimethylaminopropyl)carbodiimide and 1-hydroxy-1H-benzotriazole. Residual excess of the unreacted reagent was tagged with 2-(perfluorooctyl)ethyl isocyanate and then removed selectively using a fluorous column-switching system placed in front of an analytical reversed-phase column. The signal of the fluorous-tagged unreacted reagent was completely absent in the resulting chromatograms; therefore, it did not interfere with the quantification of each acid especially those eluted before 20 min. The detection limits (S/N = 3) for the examined acids were in the range from 4.0 to 22 fmol per injection. We have applied this method to comparative analysis of highly polar carboxylic acids in urine samples obtained from diabetes mellitus type-II model mice and their control.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Ácidos Carboxílicos
/
Cromatografia Líquida
Tipo de estudo:
Evaluation_studies
Limite:
Animals
/
Humans
/
Male
Idioma:
En
Ano de publicação:
2013
Tipo de documento:
Article