Heat bump on a monochromator crystal measured with X-ray grating interferometry.
Rutishauser, Simon; Rack, Alexander; Weitkamp, Timm; Kayser, Yves; David, Christian; Macrander, Albert T.
Afiliação
  • Rutishauser S; Paul Scherrer Institut, 5232 Villigen PSI, Switzerland. simon.rutishauser@psi.ch
J Synchrotron Radiat ; 20(Pt 2): 300-5, 2013 Mar.
Article em En | MEDLINE | ID: mdl-23412487
ABSTRACT
Deformation of the first crystal of an X-ray monochromator under the heat load of a high-power beam, commonly referred to as `heat bump', is a challenge frequently faced at synchrotron beamlines. Here, quantitative measurements of the deformations of an externally water-cooled silicon (111) double-crystal monochromator tuned to a photon energy of 17.6 keV are reported. These measurements were made using two-dimensional hard X-ray grating interferometry, a technique that enables in situ at-wavelength wavefront investigations with high angular sensitivity. The observed crystal deformations were of the order of 100 nm in the meridional and 5 nm in the sagittal direction, which lead to wavefront slope errors of up to 4 µrad in the meridional and a few hundred nanoradians in the sagittal direction.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Ano de publicação: 2013 Tipo de documento: Article