Sensitivity of assays for the detection of HPA-1a antibodies: results of an international workshop demonstrating the impact of cation chelation from integrin αIIbß3 on three widely used assays.
Allen, D L; Metcalfe, P; Kaplan, C; Kekomaki, R; de Haas, M; Yusuf, R; Ouwehand, W H.
Afiliação
  • Allen DL; Nuffield Department of Clinical Laboratory Sciences, University of Oxford, Oxford, UK. dave.allen@ndcls.ox.ac.uk
Vox Sang ; 105(2): 167-73, 2013 Aug.
Article em En | MEDLINE | ID: mdl-23662600
ABSTRACT
BACKGROUND AND

OBJECTIVES:

HPA-1a antibodies account for 70-80% of cases of fetal-neonatal alloimmune thrombocytopenia (FNAIT) in Caucasians. However, numerous workshops have demonstrated variability in their detection. We recently showed that exposure of αIIbß3 to ethylene diamine tetraacetic acid (EDTA) affected binding of many anti-αIIbß3 monoclonal, and HPA-1a allo-, antibodies; this adversely affected sensitivity of the monoclonal antibody-specific immobilization of platelet antigens (MAIPA) assay and indirect platelet immunofluorescence test (PIFT). This study presents results from an international workshop studying the impact of cation chelation on HPA-1a antibody detection in routine diagnostic laboratories. MATERIALS AND

METHODS:

Serum and EDTA-anticoagulated plasma samples containing anti-HPA-1a were distributed to 39 laboratories. Participants were asked to detect and identify any HPA antibodies present.

RESULTS:

2/39 (5.1%) participants were able to detect and identify anti-HPA-1a in the serum, but not in the plasma sample. EDTA plasma reduced MAIPA assay sensitivity by ≥ 20% in 17/24 (70.8%) laboratories and by ≥ 50% in 9/24 (37.5%) when using HPA-1a1a platelets (mean 27.7%, range 0-85.1%); when using HPA-1a1b platelets 3/4 (75%), participants reported ≥ 50% loss of sensitivity (mean 65.6%, range 0-96.6%). A small but significant increase in optical densities was observed in antigen capture ELISA assays when using plasma (mean difference 0.081, P < 0.01). Insufficient PIFT data were returned to draw firm conclusions.

CONCLUSION:

Use of EDTA plasma significantly affects the sensitivity of the MAIPA assay and can affect detection of even potent, FNAIT-causing examples of anti-HPA-1a. These data highlight the importance of use of αIIbß3 in an appropriate conformation for the sensitive detection of anti-HPA-1a.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Quelantes / Antígenos de Plaquetas Humanas / Ácido Edético / Integrina alfa2 / Integrina beta3 / Trombocitopenia Neonatal Aloimune / Isoanticorpos Tipo de estudo: Diagnostic_studies Limite: Female / Humans / Male / Newborn Idioma: En Ano de publicação: 2013 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Quelantes / Antígenos de Plaquetas Humanas / Ácido Edético / Integrina alfa2 / Integrina beta3 / Trombocitopenia Neonatal Aloimune / Isoanticorpos Tipo de estudo: Diagnostic_studies Limite: Female / Humans / Male / Newborn Idioma: En Ano de publicação: 2013 Tipo de documento: Article