Obstruction of superior vena cava flow during transcatheter atrial septal defect closure with the Atriasept ASD occluder.
Kahya Eren, Nihan; Kocabas, Ugur; Nazli, Cem; Ergene, Oktay.
Afiliação
  • Kahya Eren N; Department of Cardiology, Izmir Atatürk Training and Research Hospital, Izmir, Turkey. nkahya77@yahoo.com
Turk Kardiyol Dern Ars ; 41(2): 141-3, 2013 Mar.
Article em En | MEDLINE | ID: mdl-23666302
ABSTRACT
In this paper, we describe a patient with a large secundum atrial septal defect ASD (26 mm) with adequate rims that were suitable for percutaneous closure. While closing this ASD, the superior vena cava (SVC) was occluded with the right atrial disc of the Atriasept ASD occluder (Cardia) and thus the device was retrieved before its release. After retrieval of this device, an Amplatzer ASD occluder was successfully implanted without disturbing the caval flow.