Comparison of visual function after bilateral implantation of inferior sector-shaped near-addition and diffractive-refractive multifocal IOLs.
Rosa, Andreia Martins; Loureiro Silva, Maria Fátima; Lobo, Conceição; Mira, Joaquim Bernardes; Farinha, Cláudia Louro; Póvoa, João Alberto; Castelo-Branco, Miguel; Murta, Joaquim Neto.
Afiliação
  • Rosa AM; From the Ophthalmology Unit (Rosa, Lobo, Mira, Farinha, Póvoa, Murta), Centro Hospitalar e Universitário de Coimbra, and the Faculty of Medicine (Rosa, Lobo, Murta) and Visual Neuroscience Laboratory (Silva, Castelo-Branco), IBILI, Faculty of Medicine, Coimbra University, Coimbra, Portugal.
J Cataract Refract Surg ; 39(11): 1653-9, 2013 Nov.
Article em En | MEDLINE | ID: mdl-24160381
ABSTRACT

PURPOSE:

To compare visual function after bilateral implantation of multifocal Lentis Mplus LS-312 (Group A) or Acrysof Restor SN6AD1 (Group B) intraocular lenses (IOLs).

SETTING:

Ophthalmology Unit, Centro Hospitalar e Universitário de Coimbra, and Visual Neuroscience Laboratory, IBILI, Faculty of Medicine, University of Coimbra, Coimbra, Portugal.

DESIGN:

Comparative case series.

METHODS:

Patients between 49 years and 76 years had bilateral cataract surgery with multifocal IOL implantation. Patients were evaluated preoperatively and 3 months postoperatively for distance, intermediate, and near visual acuities; static photopic and mesopic contrast sensitivity; and visual acuity under a glare source using the Metrovision contrast sensitivity platform. Color vision was evaluated with the Cambridge Colour Test.

RESULTS:

Group A comprised 56 eyes and Group B, 44 eyes. Visual and refractive results were comparable between the 2 groups. Photopic contrast sensitivity was significantly better in Group B at intermediate (2.2 cycles per degree [cpd] and 3.4 cpd) and high (7.1 cpd and 23.6 cpd) spatial frequencies. Under low mesopic conditions (0.08 candelas/m(2)), differences were significant at 1.1 cpd and 2.2 cpd spatial frequencies. There were no differences in visual acuity under a glare source or in color vision.

CONCLUSIONS:

Both IOLs provided good distance, intermediate, and near visual acuities. Visual acuity under a glare source and color vision were similar in the 2 groups. However, photopic and low mesopic contrast sensitivities were better in Group B, particularly for intermediate spatial frequencies, which are important for night driving.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Sensibilidades de Contraste / Acuidade Visual / Ofuscação / Pseudofacia / Implante de Lente Intraocular / Visão de Cores / Lentes Intraoculares Tipo de estudo: Observational_studies Limite: Aged / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2013 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Sensibilidades de Contraste / Acuidade Visual / Ofuscação / Pseudofacia / Implante de Lente Intraocular / Visão de Cores / Lentes Intraoculares Tipo de estudo: Observational_studies Limite: Aged / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2013 Tipo de documento: Article