CT-guided, minimally invasive, postmortem needle biopsy using the B-Rob II needle-positioning robot.
J Forensic Sci
; 59(2): 517-21, 2014 Mar.
Article
em En
| MEDLINE
| ID: mdl-24313538
ABSTRACT
CT-guided, minimally invasive needle biopsy techniques can be used to retrieve tissue or body fluid samples for histologic tissue diagnoses in forensic investigations. The purpose of this study was an evaluation of CT-guided needle-positioning robot B-Rob II. To operate under CT guidance, we adopted the B-Rob II robotic needle placement system and its workflow. The accuracy and speed of the procedure were tested on a gelatin phantom in a series of 21 biopsies. We achieved an average needle placement accuracy of 1.8 mm (±1.1 mm) using robotic assistance. The procedure required an average of 2 min 21 s. The needle placement accuracy for minimally invasive needle biopsies using the B-Rob II biopsy robot is sufficiently accurate and fast for forensic postmortem examinations of focal organ changes. Further tests will be performed to test the feasibility of the robot for performing biopsies of focal organ changes in human bodies.