Nexfin noninvasive continuous hemodynamic monitoring: validation against continuous pulse contour and intermittent transpulmonary thermodilution derived cardiac output in critically ill patients.
Ameloot, Koen; Van De Vijver, Katrijn; Broch, Ole; Van Regenmortel, Niels; De Laet, Inneke; Schoonheydt, Karen; Dits, Hilde; Bein, Berthold; Malbrain, Manu L N G.
Afiliação
  • Ameloot K; Department of Intensive Care Medicine, Ziekenhuis Netwerk Antwerpen, ZNA Stuivenberg, Lange Beeldekensstraat 267, 2060 Antwerp, Belgium.
ScientificWorldJournal ; 2013: 519080, 2013.
Article em En | MEDLINE | ID: mdl-24319373
ABSTRACT

INTRODUCTION:

Nexfin (Bmeye, Amsterdam, Netherlands) is a noninvasive cardiac output (CO) monitor based on finger arterial pulse contour analysis. The aim of this study was to validate Nexfin CO (NexCO) against thermodilution (TDCO) and pulse contour CO (CCO) by PiCCO (Pulsion Medical Systems, Munich, Germany). PATIENTS AND

METHODS:

In a mix of critically ill patients (n = 45), NexCO and CCO were measured continuously and recorded at 2-hour intervals during the 8-hour study period. TDCO was measured at 0-4-8 hrs.

RESULTS:

NexCO showed a moderate to good (significant) correlation with TDCO (R (2) 0.68, P < 0.001) and CCO (R (2) 0.71, P < 0.001). Bland and Altman analysis comparing NexCO with TDCO revealed a bias (± limits of agreement, LA) of 0.4 ± 2.32 L/min (with 36% error) while analysis comparing NexCO with CCO showed a bias (±LA) of 0.2 ± 2.32 L/min (37% error). NexCO is able to follow changes in TDCO and CCO during the same time interval (level of concordance 89.3% and 81%). Finally, polar plot analysis showed that trending capabilities were acceptable when changes in NexCO (ΔNexCO) were compared to ΔTDCO and ΔCCO (resp., 89% and 88.9% of changes were within the level of 10% limits of agreement).

CONCLUSION:

we found a moderate to good correlation between CO measurements obtained with Nexfin and PiCCO.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Débito Cardíaco / Hemodinâmica / Monitorização Fisiológica Tipo de estudo: Observational_studies Limite: Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2013 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Débito Cardíaco / Hemodinâmica / Monitorização Fisiológica Tipo de estudo: Observational_studies Limite: Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2013 Tipo de documento: Article