Diagnostic accuracy of digitized chest X-rays using consumer-grade color displays for low-cost teleradiology services: a multireader-multicase comparison.
Salazar, Antonio J; Aguirre, Diego A; Ocampo, Juliana; Diaz, Xavier A; Camacho, Juan C.
Afiliação
  • Salazar AJ; 1 Department of Electrical and Electronic Engineering, University of Los Andes , Bogotá, Colombia .
Telemed J E Health ; 20(4): 304-11, 2014 Apr.
Article em En | MEDLINE | ID: mdl-24506568
ABSTRACT

INTRODUCTION:

In teleradiology services and in hospitals, the extensive use of visualization displays requires affordable devices. The purpose of this study was to compare three differently priced displays (a medical-grade grayscale display and two consumer-grade color displays) for image visualization of digitized chest X-rays. MATERIALS AND

METHODS:

The evaluated conditions were interstitial opacities, pneumothorax, and nodules using computed tomography as the gold standard. The comparison was accomplished in terms of receiver operating characteristic (ROC) curves, the diagnostic power measured as the area under ROC curves, accuracy in conditions classification, and main factors affecting accuracy, in a factorial study with 76 cases and six radiologists.

RESULTS:

The ROC curves for all of the displays and pathologies had similar shapes and no differences in diagnostic power. The proportion of cases correctly classified for each display was greater than 71.9%. The correctness proportions of the three displays were different (p<0.05) only for interstitial opacities. The evaluation of the main factors affecting these proportions revealed that the display factor was not significant for either nodule size or pneumothorax size (p>0.05).

CONCLUSIONS:

Although the image quality variables showed differences in the radiologists' perceptions of the image quality of the three displays, significant differences in the accuracy did not occur. The main effect on the variability of the proportions of correctly classified cases did not come from the display factor. This study confirms previous findings that medical-grade displays could be replaced by consumer-grade color displays with the same image quality.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Apresentação de Dados / Radiografia Torácica / Tomografia Computadorizada por Raios X / Telerradiologia Tipo de estudo: Diagnostic_studies / Health_economic_evaluation Limite: Humans Idioma: En Ano de publicação: 2014 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Apresentação de Dados / Radiografia Torácica / Tomografia Computadorizada por Raios X / Telerradiologia Tipo de estudo: Diagnostic_studies / Health_economic_evaluation Limite: Humans Idioma: En Ano de publicação: 2014 Tipo de documento: Article