Development of a sensitive LC/MS/MS method for vitamin D metabolites: 1,25 Dihydroxyvitamin D2&3 measurement using a novel derivatization agent.
Hedman, Curtis J; Wiebe, Donald A; Dey, Subhakar; Plath, Josh; Kemnitz, Joseph W; Ziegler, Toni E.
Afiliação
  • Hedman CJ; Institute of Clinical and Translational Research, University of Wisconsin, Madison, WI, USA; Wisconsin National Primate Research Center, University of Wisconsin, Madison, WI, USA; Wisconsin State Laboratory of Hygiene, University of Wisconsin, Madison, WI, USA.
  • Wiebe DA; Department of Pathology and Laboratory Medicine, University of Wisconsin, Madison, WI, USA.
  • Dey S; AB SCIEX, Chemistry and Consumables R and D, Framingham, MA, USA.
  • Plath J; AB SCIEX, Chemistry and Consumables R and D, Framingham, MA, USA.
  • Kemnitz JW; Institute of Clinical and Translational Research, University of Wisconsin, Madison, WI, USA; Wisconsin National Primate Research Center, University of Wisconsin, Madison, WI, USA; Department of Cell and Regenerative Biology, University of Wisconsin, Madison, WI, USA.
  • Ziegler TE; Institute of Clinical and Translational Research, University of Wisconsin, Madison, WI, USA; Wisconsin National Primate Research Center, University of Wisconsin, Madison, WI, USA. Electronic address: ziegler@primate.wisc.edu.
Article em En | MEDLINE | ID: mdl-24576767
ABSTRACT
Active vitamin D metabolites 1,25-dihydroxyvitamin D2 [1,25-(OH)2-D2; derived from ergocalciferol] and D3 [1,25-(OH)2-D3; derived from cholecalciferol] are found in low levels in the circulation and require a very sensitive method for measurement. Radioimmunoassay (RIA) has been the method of choice, but it lacks the specificity needed to distinguish between 1,25-(OH)2-D2 and -D3, whereas liquid chromatography-tandem mass spectrometry (LC/MS/MS) methods have the advantage of high specificity and sensitivity. Here, we compare a new derivative for ionizing 1,25-(OH)2-D to enhance the signal and provide the most sensitive assay for measuring vitamin D. We used the Amplifex diene method of derivatizing prior to LC/MS/MS and compared it to the standard RIA method and the 4-phenyl-1,2,4-triazole-3,5-dione (PTAD) method of derivatizing prior to LC/MS/MS. In the evaluation of 20 human serum samples, all methods correlated strongly across the upper levels of the standard 1,25-(OH)2-D2 and -D3 ranges (Amplifex and RIA, pc=0.97; Amplifex and PTAD, pc=0.96) but less strongly on the lower levels of the standard range (Amplifex and RIA, pc=0.81; Amplifex and PTAD, pc=0.65) suggesting differences in the sensitivities between the assays. The Amplifex method was determined to be more sensitive than the PTAD method, as peak areas were significantly higher for the Amplifex method and provided for a 10 fold higher signal-to-noise ratio than PTAD. Therefore, the Amplifex LC/MS/MS method is the most sensitive and specific method available for measuring 1,25-(OH)2-D2 and -D3 while using the smallest sample volume.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Calcitriol / Ergocalciferóis / Cromatografia Líquida / Espectrometria de Massas em Tandem Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Humans Idioma: En Ano de publicação: 2014 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Calcitriol / Ergocalciferóis / Cromatografia Líquida / Espectrometria de Massas em Tandem Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Humans Idioma: En Ano de publicação: 2014 Tipo de documento: Article