Your browser doesn't support javascript.
loading
Counting rate enhancements in superconducting nanowire single-photon detectors with improved readout circuits.
Opt Lett ; 39(7): 1869-72, 2014 Apr 01.
Article em En | MEDLINE | ID: mdl-24686626
ABSTRACT
Counting rates of superconducting nanowire single-photon detectors are usually estimated at hundreds of MHz by their kinetic-inductive reset time. This maximum is also limited by capacitor coupling effects in conventional readout circuits. In this Letter, we design and demonstrate an improved readout circuit that reduces the reset time and removes circuit limits. The counting rate at the 3 dB compression point is increased by four times for a large active area detector. We also discuss nonlinear dependences of the counting rate on the incident continuous-wave optical power and give a numerical model to explain our observations.

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Ano de publicação: 2014 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Ano de publicação: 2014 Tipo de documento: Article