Your browser doesn't support javascript.
loading
The additional value of three-dimensional transesophageal echocardiography in complex aortic prosthetic heart valve endocarditis.
Tanis, Wilco; Teske, Arco J; van Herwerden, Lex A; Chamuleau, Steven; Meijboom, Folkert; Budde, Ricardo P J; Cramer, Maarten-Jan.
Afiliação
  • Tanis W; Department of Cardiology, University Medical Center Utrecht, Utrecht, The Netherlands; Department of Cardiology, Haga Teaching Hospital, The Hague, The Netherlands.
Echocardiography ; 32(1): 114-25, 2015 Jan.
Article em En | MEDLINE | ID: mdl-24724568
ABSTRACT

BACKGROUND:

Two-dimensional transthoracic and transesophageal echocardiography (2DTTE and 2DTEE) may fail to detect signs of prosthetic heart valve (PHV) endocarditis due to acoustic shadowing. Three-dimensional (3D) TEE may have additional value; however, data are scarce. This study was performed to investigate the additional value of 3DTEE for the detection of aortic PHV endocarditis and the extent of the disease process.

METHODS:

Retrospective analysis of complex aortic PHV endocarditis cases that underwent 2DTTE, 2DTEE, and 3DTEE before surgery. Echocardiograms were individually assessed by 2 cardiologists blinded for the outcome. Surgical and pathological inspection served as the reference standard for vegetations and peri-annular extensions (abscesses/mycotic aneurysms). To determine if the proximal coronary arteries were involved in the inflammatory process as well, computed tomography angiography findings were added to reference standard.

RESULTS:

Fifteen aortic PHV endocarditis cases were identified. According to the reference standard, all 15 cases had peri-annular extensions, 13 of which had a close relationship with the proximal right and/or left coronary artery. In 6 of 15 patients, a vegetation was present. Combined 2DTTE/TEE missed 1/6 vegetations and 1/15 peri-annular extensions. After addition of 3DTEE, all vegetations (6/6) and peri-annular extensions (15/15) were detected, without adding false positives. Compared to 2DTEE, in 3/15 cases, 3DTEE resulted in better delineation of the anatomical relationship of the proximal coronary arteries to the peri-annular extensions. As a result, 3DTEE had an additional value in 5/15 cases.

CONCLUSION:

In complex aortic, PHV endocarditis 3DTEE may have additional value compared to 2D echocardiography.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Valva Aórtica / Próteses Valvulares Cardíacas / Infecções Relacionadas à Prótese / Endocardite Tipo de estudo: Diagnostic_studies / Evaluation_studies / Observational_studies / Prognostic_studies Limite: Aged / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2015 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Valva Aórtica / Próteses Valvulares Cardíacas / Infecções Relacionadas à Prótese / Endocardite Tipo de estudo: Diagnostic_studies / Evaluation_studies / Observational_studies / Prognostic_studies Limite: Aged / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2015 Tipo de documento: Article