Your browser doesn't support javascript.
loading
Corrosion resistance, surface evaluation, and geometric design comparison of five self-expanding nitinol stents used in clinical practice.
Morrison, Jasper W; Pelletier, Matthew H; Rives, Alain; Walsh, William R; Yang, Jia-Lin; Varcoe, Ramon L.
Afiliação
  • Morrison JW; 1 Department of Surgery and.
J Endovasc Ther ; 21(2): 230-9, 2014 Apr.
Article em En | MEDLINE | ID: mdl-24754282
ABSTRACT

PURPOSE:

To investigate the corrosion resistance properties of 5 commercially available nitinol stents used to treat peripheral artery disease and compare their surface quality, elemental composition, and geometrical design.

METHODS:

Samples of 5 different designs of nitinol peripheral stents [LifeStent (n=4), Philon (n=6), Epic (n=6), S.M.A.R.T. Control (n=7), and Complete SE (n=7)] were examined using stereomicroscopy, environmental scanning electron microscopy, and x-ray photoelectron spectroscopy. Corrosion resistance testing was performed in accordance with ASTM International Standard F2129-08.

RESULTS:

Thirteen (43%) of 30 stents corroded during this experiment. Stent fracture was observed in 12 (92%) of these corroded stents. Mean breakdown potentials ranged from 517 to 835 mV (vs. Ag/AgCl) for the Philon, Complete SE, S.M.A.R.T. Control, Epic, and LifeStent models from lowest to highest. A statistically significant difference in breakdown potential was observed between the LifeStent vs. Philon stents (835 vs. 517 mV, p=0.01) and Epic vs. Philon stents (833 vs. 517 mV, p=0.03). Stents with lower breakdown potential and relative breakdown potentials were associated with a higher fracture frequency (Spearman correlation coefficient -0.44, p=0.015 and -0.869, p<0.01, respectively).

CONCLUSION:

In this in vitro study, corrosion led independently to stent fracture. There is a significant association between lower mean breakdown/relative breakdown potentials and stent fracture.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Desenho de Prótese / Falha de Prótese / Stents / Ligas / Doença Arterial Periférica / Procedimentos Endovasculares Tipo de estudo: Evaluation_studies Limite: Humans Idioma: En Ano de publicação: 2014 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Desenho de Prótese / Falha de Prótese / Stents / Ligas / Doença Arterial Periférica / Procedimentos Endovasculares Tipo de estudo: Evaluation_studies Limite: Humans Idioma: En Ano de publicação: 2014 Tipo de documento: Article