ABSTRACT
We evaluated the feasibility, safety, and efficacy of sequential pulse countershock (SqCS) delivered solely through two endocardial catheters for the termination of ventricular tachycardia (VT) and fibrillation (VF) in patients undergoing electrophysiology studies (EPS). Thirty-four patients (31 men, 3 women) with a mean age of 56.8 +/- 10.1 years were studied. Etiology of VT/VF was ischemic heart disease (n = 26), cardiomyopathy (4) repaired tetralogy of Fallot (n = 1), heart transplant (n = 1), and no identifiable heart disease (n = 2). Catheters were positioned successfully in 29 patients. These were positioned in the right ventricular apex (RVA) and the coronary sinus (CS), respectively. The RVA electrode served as the common cathode for both pulses. The two electrodes located near the right atrium/superior vena cava junction served as anode for pulse 1 while the distal CS electrodes served as anode for pulse 2. Twenty-nine induced VT episodes with cycle length (CL) 220-370 msec were treated. SqCS successfully terminated 15 VT (100-500V) while 14 were accelerated or degenerated to VF. VTCL was longer in successful SqCS episodes than in those that were accelerated (285 +/- 17.3 vs 245 +/- 30.8 msec, P less than .003). Of 26 VF episodes, 21 were terminated with SqCS (500-900V) and 5 were terminated by transthoracic rescue shocks. On 2 occasions, failure to defibrillate was attributable to poor catheter position at the time of shock. No complications occurred. We conclude that SqCS delivered solely between endocardial catheter electrodes is feasible and effective using energy doses within the range of existing implantable cardioverter defibrillators.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Cardioversão Elétrica Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 1989 Tipo de documento: Article
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Cardioversão Elétrica Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 1989 Tipo de documento: Article