Your browser doesn't support javascript.
loading
Left atrial thrombus after complete left atrial appendage exclusion with LARIAT device.
Koranne, Ketan P; Fernando, Rajeev R; Laing, Susan T.
Afiliação
  • Koranne KP; Division of Cadiovascular Medicine, Department of Internal Medicine, University of Texas Health Science Center at Houston, Houston, Texas.
Catheter Cardiovasc Interv ; 85(2): E54-7, 2015 Feb 01.
Article em En | MEDLINE | ID: mdl-24846552
ABSTRACT
A 74-year-old woman with paroxysmal atrial fibrillation underwent left atrial appendage (LAA) exclusion with a LARIAT snare device. Transesophageal echocardiogram one month later demonstrated a left atrial thrombus at orifice of the completely occluded LAA that subsequently resolved with two months of anticoagulation. This case highlights that LAA ligation with LARIAT device continues to pose a risk for left atrial thrombosis in the immediate post-operative period. It also emphasizes the need for further evidence regarding antiplatelet and anticoagulation therapy in these patients. © 2014 Wiley Periodicals, Inc.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Fibrilação Atrial / Trombose / Cateterismo Cardíaco / Apêndice Atrial / Cardiopatias Tipo de estudo: Diagnostic_studies Limite: Aged / Female / Humans Idioma: En Ano de publicação: 2015 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Fibrilação Atrial / Trombose / Cateterismo Cardíaco / Apêndice Atrial / Cardiopatias Tipo de estudo: Diagnostic_studies Limite: Aged / Female / Humans Idioma: En Ano de publicação: 2015 Tipo de documento: Article