Ultra-low-dose computed tomographic angiography with model-based iterative reconstruction compared with standard-dose imaging after endovascular aneurysm repair: a prospective pilot study.
Naidu, Sailen G; Kriegshauser, J Scott; Paden, Robert G; He, Miao; Wu, Qing; Hara, Amy K.
Afiliação
  • Naidu SG; Department of Radiology, Mayo Clinic Hospital, 5777 E Mayo Blvd, 85054, Phoenix, AZ, USA, naidu.sailen@mayo.edu.
Abdom Imaging ; 39(6): 1297-303, 2014 Dec.
Article em En | MEDLINE | ID: mdl-24859850
ABSTRACT

PURPOSE:

An ultra-low-dose radiation protocol reconstructed with model-based iterative reconstruction was compared with our standard-dose protocol.

METHODS:

This prospective study evaluated 20 men undergoing surveillance-enhanced computed tomography after endovascular aneurysm repair. All patients underwent standard-dose and ultra-low-dose venous phase imaging; images were compared after reconstruction with filtered back projection, adaptive statistical iterative reconstruction, and model-based iterative reconstruction. Objective measures of aortic contrast attenuation and image noise were averaged. Images were subjectively assessed (1 = worst, 5 = best) for diagnostic confidence, image noise, and vessel sharpness. Aneurysm sac diameter and endoleak detection were compared.

RESULTS:

Quantitative image noise was 26% less with ultra-low-dose model-based iterative reconstruction than with standard-dose adaptive statistical iterative reconstruction and 58% less than with ultra-low-dose adaptive statistical iterative reconstruction. Average subjective noise scores were not different between ultra-low-dose model-based iterative reconstruction and standard-dose adaptive statistical iterative reconstruction (3.8 vs. 4.0, P = .25). Subjective scores for diagnostic confidence were better with standard-dose adaptive statistical iterative reconstruction than with ultra-low-dose model-based iterative reconstruction (4.4 vs. 4.0, P = .002). Vessel sharpness was decreased with ultra-low-dose model-based iterative reconstruction compared with standard-dose adaptive statistical iterative reconstruction (3.3 vs. 4.1, P < .0001). Ultra-low-dose model-based iterative reconstruction and standard-dose adaptive statistical iterative reconstruction aneurysm sac diameters were not significantly different (4.9 vs. 4.9 cm); concordance for the presence of endoleak was 100% (P < .001).

CONCLUSION:

Compared with a standard-dose technique, an ultra-low-dose model-based iterative reconstruction protocol provides comparable image quality and diagnostic assessment at a 73% lower radiation dose.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Aorta Abdominal / Complicações Pós-Operatórias / Doses de Radiação / Interpretação de Imagem Radiográfica Assistida por Computador / Aneurisma da Aorta Abdominal / Tomografia Computadorizada Multidetectores Tipo de estudo: Diagnostic_studies / Guideline / Observational_studies / Prognostic_studies Limite: Aged / Aged80 / Humans / Male / Middle aged Idioma: En Ano de publicação: 2014 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Aorta Abdominal / Complicações Pós-Operatórias / Doses de Radiação / Interpretação de Imagem Radiográfica Assistida por Computador / Aneurisma da Aorta Abdominal / Tomografia Computadorizada Multidetectores Tipo de estudo: Diagnostic_studies / Guideline / Observational_studies / Prognostic_studies Limite: Aged / Aged80 / Humans / Male / Middle aged Idioma: En Ano de publicação: 2014 Tipo de documento: Article