Effect of cricoid pressure on placement of the I-gel™ : a randomised study.
Anaesthesia
; 69(8): 878-82, 2014 Aug.
Article
em En
| MEDLINE
| ID: mdl-24866121
ABSTRACT
We studied 40 adult patients to see if cricoid pressure affected placement of the I-gel(™). In a randomised crossover design, the i-gel was placed with and without cricoid pressure, and we compared the success rate of adequate ventilation through the i-gel, time to placement and the rate of optimal position of the device between the two circumstances. Cricoid pressure significantly decreased the success rate of adequate ventilation through the i-gel (40 vs 34 patients) (p = 0.041, 95% CI for difference 4-26%), and significantly decreased the rate of the optimal position (39 vs 17 patients) (p < 0.001). The time to achieve adequate ventilation was significantly longer (p < 0.001) with cricoid pressure than without (median difference 8 s; 95% CI for median difference 3-12 s). Cricoid pressure significantly decreases the success rate of ventilation through the i-gel, but the success rate of ventilation through the i-gel is reasonably high.