Forces and moments generated by removable thermoplastic aligners: incisor torque, premolar derotation, and molar distalization.
Simon, Mareike; Keilig, Ludger; Schwarze, Jörg; Jung, Britta A; Bourauel, Christoph.
Afiliação
  • Simon M; Orthodontist, Department of Orthodontics, University Hospital of Freiburg, Freiburg, Germany; doctoral candidate, Department of Oral Technology, School of Dentistry, University of Bonn, Bonn, Germany. Electronic address: mareike.simon@uniklinik-freiburg.de.
  • Keilig L; Senior scientist, Department of Oral Technology, School of Dentistry, University of Bonn, Bonn, Germany.
  • Schwarze J; Private practice, Cologne, Germany.
  • Jung BA; Professor and head of clinic, Department of Orthodontics, University Hospital of Freiburg, Freiburg, Germany.
  • Bourauel C; Professor and head, Orthodontics Department, School of Dentistry, University of Bonn, Bonn, Germany.
Am J Orthod Dentofacial Orthop ; 145(6): 728-36, 2014 Jun.
Article em En | MEDLINE | ID: mdl-24880843
INTRODUCTION: The exact force systems as well as their progressions generated by removable thermoplastic appliances have not been investigated. Thus, the purposes of this experimental study were to quantify the forces and moments delivered by a single aligner and a series of aligners (Invisalign; Align Technology, Santa Clara, Calif) and to investigate the influence of attachments and power ridges on the force transfer. METHODS: We studied 970 aligners of the Invisalign system (60 series of aligners). The aligners came from 30 consecutive patients, of which 3 tooth movements (incisor torque, premolar derotation, molar distalization) with 20 movements each were analyzed. The 3 movement groups were subdivided so that 10 movements were supported with an attachment and 10 were not. The patients' ClinCheck (Align Technology, Santa Clara, Calif) was planned so that the movements to be investigated were performed in isolation in the respective quadrant. Resin replicas of the patients' intraoral situation before the start of the investigated movement were taken and mounted in a biomechanical measurement system. An aligner was put on the model, the force systems were measured, and the calculated movements were experimentally performed until no further forces or moments were generated. Subsequently, the next aligners were installed, and the measurements were repeated. RESULTS: The initial mean moments were about 7.3 N·mm for maxillary incisor torque and about 1.0 N for distalization. Significant differences in the generated moments were measured in the premolar derotation group, whether they were supported with an attachment (8.8 N·mm) or not (1.2 N·mm). All measurements showed an exponential force change. CONCLUSIONS: Apart from a few maximal initial force systems, the forces and moments generated by aligners of the Invisalign system are within the range of orthodontic forces. The force change is exponential while a patient is wearing removable thermoplastic appliances.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Aparelhos Ortodônticos Removíveis / Técnicas de Movimentação Dentária / Dente Pré-Molar / Desenho de Aparelho Ortodôntico / Incisivo / Dente Molar Tipo de estudo: Clinical_trials Limite: Adolescent / Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2014 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Aparelhos Ortodônticos Removíveis / Técnicas de Movimentação Dentária / Dente Pré-Molar / Desenho de Aparelho Ortodôntico / Incisivo / Dente Molar Tipo de estudo: Clinical_trials Limite: Adolescent / Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2014 Tipo de documento: Article