Your browser doesn't support javascript.
loading
Implantation of the new Nit-Occlud PDA-R device in children below 10 kilogram.
Moysich, Axel; Happel, Christoph M; Laser, Kai T; Kececioglu, Deniz; Haas, Nikolaus A.
Afiliação
  • Moysich A; Department of Congenital Heart Defects, Heart and Diabetes Centre North Rhine Westphalia, Ruhr University Bochum, Germany.
  • Happel CM; Department of Congenital Heart Defects, Heart and Diabetes Centre North Rhine Westphalia, Ruhr University Bochum, Germany.
  • Laser KT; Department of Congenital Heart Defects, Heart and Diabetes Centre North Rhine Westphalia, Ruhr University Bochum, Germany.
  • Kececioglu D; Department of Congenital Heart Defects, Heart and Diabetes Centre North Rhine Westphalia, Ruhr University Bochum, Germany.
  • Haas NA; Department of Congenital Heart Defects, Heart and Diabetes Centre North Rhine Westphalia, Ruhr University Bochum, Germany.
Catheter Cardiovasc Interv ; 85(7): 1203-8, 2015 Jun.
Article em En | MEDLINE | ID: mdl-25258039
ABSTRACT

BACKGROUND:

Interventional closure of patent ductus arteriosus (PDA) has become a common and safe procedure in most pediatric cath labs. Interventional treatment of PDAs still remains a challenge in those children with low body weight and a large PDA. The Nit-Occlud PDA-R® device was developed and especially designed for large PDAs. We report our most recent experience in children with a body weight lower than 10 kg. MATERIALS AND

METHODS:

The PDA-R® device was used in seven children (age 1-10, median 6 months) with a body weight from 4.1 to 9.7 kg (median 5.9 kg) ductal length was 12 mm (median), with a large ampulla (median 9 mm) which exceeded the diameter of the aorta (median 6 mm) and large diameter (median minimal diameter 4 mm). In six cases, the Nit-Occlud PDA-R was selected with an aortic disc of 12 mm and in one case an occluder with an aortic disc of 14 mm.

RESULTS:

Occlusion of the PDA was documented by angiography and/or echocardiography in all cases. At a mean follow-up of 21.4 months, no flow obstruction to the left or right pulmonary artery or new onset coarctation of the aorta was noted.

CONCLUSIONS:

The Nit-Occlud PDA-R® device is suitable in children with a body weight below 10 kg when a relative large PDA is present.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Peso Corporal / Cateterismo Cardíaco / Permeabilidade do Canal Arterial / Dispositivo para Oclusão Septal Tipo de estudo: Diagnostic_studies / Etiology_studies / Observational_studies / Risk_factors_studies Limite: Female / Humans / Infant / Male Idioma: En Ano de publicação: 2015 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Peso Corporal / Cateterismo Cardíaco / Permeabilidade do Canal Arterial / Dispositivo para Oclusão Septal Tipo de estudo: Diagnostic_studies / Etiology_studies / Observational_studies / Risk_factors_studies Limite: Female / Humans / Infant / Male Idioma: En Ano de publicação: 2015 Tipo de documento: Article