Pre-Column Derivatization HPLC Procedure for the Quantitation of Aluminium Chlorohydrate in Antiperspirant Creams Using Quercetin as Chromogenic Reagent.
Kalogria, Eleni; Varvaresou, Athanasia; Papageorgiou, Spyridon; Protopapa, Evaggelia; Tsaknis, Ioannis; Matikas, Alexios; Panderi, Irene.
Afiliação
  • Kalogria E; Division of Pharmaceutical Chemistry, Department of Pharmacy, University of Athens, Panepistimiopolis, Zografou, 157 71 Athens, Greece.
  • Varvaresou A; Department of Aesthetics and Cosmetology, The School of Health and Caring Professions, Technological Education Institution of Athens, Ag. Spyridonos str., 122 10 Athens, Greece.
  • Papageorgiou S; Department of Aesthetics and Cosmetology, The School of Health and Caring Professions, Technological Education Institution of Athens, Ag. Spyridonos str., 122 10 Athens, Greece.
  • Protopapa E; Department of Aesthetics and Cosmetology, The School of Health and Caring Professions, Technological Education Institution of Athens, Ag. Spyridonos str., 122 10 Athens, Greece.
  • Tsaknis I; Department of Aesthetics and Cosmetology, The School of Health and Caring Professions, Technological Education Institution of Athens, Ag. Spyridonos str., 122 10 Athens, Greece.
  • Matikas A; Medical Oncology Department, University Hospital of Heraklion, Mainz, Greece.
  • Panderi I; Division of Pharmaceutical Chemistry, Department of Pharmacy, University of Athens, Panepistimiopolis, Zografou, 157 71 Athens, Greece.
Chromatographia ; 77(19-20): 1275-1281, 2014.
Article em En | MEDLINE | ID: mdl-25278619
ABSTRACT
This article describes the development and validation of a selective high-performance liquid chromatography method that allows, after liquid-liquid extraction and pre-column derivatization reaction with quercetin, the quantification of aluminium chlorohydrate in antiperspirant creams. Chromatographic separation was achieved on an XTerra MS C18 analytical column (150 × 3.0 mm i.d., particle size 5 µm) using a mobile phase of acetonitrilewater (1585, v/v) containing 0.08 % trifluoroacetic acid at a flow rate of 0.30 mL min-1. Ultraviolet spectrophotometric detection at 415 nm was used. The assay was linear over a concentration range of 3.7-30.6 µg mL-1 for aluminium with a limit of quantitation of 3.74 µg mL-1. Quality control samples (4.4, 17.1 and 30.6 µg mL-1) in five replicates from five different runs of analysis demonstrated intra-assay precision (% coefficient of variation <3.8 %), inter-assay precision (% coefficient of variation <5.4 %) and an overall accuracy (% recovery) between 96 and 101 %. The method was used to quantify aluminium in antiperspirant creams containing 11.0, 13.0 and 16.0 % (w/w) aluminium chlorohydrate, respectively.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Ano de publicação: 2014 Tipo de documento: Article