Efficiency of gas transfer in venovenous extracorporeal membrane oxygenation: analysis of 317 cases with four different ECMO systems.
Lehle, Karla; Philipp, Alois; Hiller, Karl-Anton; Zeman, Florian; Buchwald, Dirk; Schmid, Christof; Dornia, Christian; Lunz, Dirk; Müller, Thomas; Lubnow, Matthias.
Afiliação
  • Lehle K; Department of Cardiothoracic Surgery, University Medical Center Regensburg, 93042, Regensburg, Germany.
Intensive Care Med ; 40(12): 1870-7, 2014 Dec.
Article em En | MEDLINE | ID: mdl-25323118
PURPOSE: Polymethylpentene membrane oxygenators used in venovenous extracorporeal membrane oxygenation (vvECMO) differ in their physical characteristics. The aim of the study was to analyze the gas transfer capability of different ECMO systems in clinical practice, as the choice of the appropriate system may be influenced by the needs of the patient. METHODS: Retrospective study on prospectively collected data of adults with severe respiratory failure requiring vvECMO support (Regensburg ECMO Registry, 2009-2013). Oxygen (O2) transfer and carbon dioxide (CO2) elimination of four different ECMO systems (PLS system, n = 163; Cardiohelp system (CH), n = 59, Maquet Cardiopulmonary, Rastatt, Germany; Hilite 7000 LT system, n = 56, Medos Medizintechnik, Stolberg, Germany; ECC.05 system, n = 39, Sorin Group, Mirandola (MO), Italy) were analyzed. RESULTS: Gas transfer depended on type of ECMO system, blood flow, and gas flow (p ≤ 0.05, each). CO2 removal is dependent on sweep gas flow and blood flow, with higher blood flow and/or gas flow eliminating more CO2 (p ≤ 0.001). CO2 elimination capacity was highest with the PLS system (p ≤ 0.001). O2 transfer at blood flow rates below 3 l/min depended on blood flow, at higher blood flow rates on blood flow and gas flow. The system with the smallest gas exchange surface (ECC.05 system) was least effective in O2 transfer, but in terms of the gas exchange surface was the most effective. CONCLUSION: Our analysis suggests that patients with severe hypoxemia and need for high flow ECMO benefit more from the PLS/CH or Hilite 7000 LT system. The ECC.05 system is advisable for patients with moderate hypoxemia and/or hypercapnia.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Oxigênio / Insuficiência Respiratória / Dióxido de Carbono / Oxigenação por Membrana Extracorpórea / Hipercapnia / Hipóxia Tipo de estudo: Etiology_studies / Observational_studies / Risk_factors_studies Limite: Adult / Aged / Female / Humans / Male / Middle aged País/Região como assunto: Europa Idioma: En Ano de publicação: 2014 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Oxigênio / Insuficiência Respiratória / Dióxido de Carbono / Oxigenação por Membrana Extracorpórea / Hipercapnia / Hipóxia Tipo de estudo: Etiology_studies / Observational_studies / Risk_factors_studies Limite: Adult / Aged / Female / Humans / Male / Middle aged País/Região como assunto: Europa Idioma: En Ano de publicação: 2014 Tipo de documento: Article