Predicting the sensitivity of the beryllium/scintillator layer neutron detector using Monte Carlo and experimental response functions.
Rev Sci Instrum
; 85(11): 11E617, 2014 Nov.
Article
em En
| MEDLINE
| ID: mdl-25430363
A methodology for obtaining empirical curves relating absolute measured scintillation light output to beta energy deposited is presented. Output signals were measured from thin plastic scintillator using NIST traceable beta and gamma sources and MCNP5 was used to model the energy deposition from each source. Combining the experimental and calculated results gives the desired empirical relationships. To validate, the sensitivity of a beryllium/scintillator-layer neutron activation detector was predicted and then exposed to a known neutron fluence from a Deuterium-Deuterium fusion plasma (DD). The predicted and the measured sensitivity were in statistical agreement.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Tipo de estudo:
Diagnostic_studies
/
Prognostic_studies
/
Risk_factors_studies
Idioma:
En
Ano de publicação:
2014
Tipo de documento:
Article