Unsupervised home use of an overnight closed-loop system over 3-4 weeks: a pooled analysis of randomized controlled studies in adults and adolescents with type 1 diabetes.
Thabit, H; Elleri, D; Leelarathna, L; Allen, J M; Lubina-Solomon, A; Stadler, M; Walkinshaw, E; Iqbal, A; Choudhary, P; Wilinska, M E; Barnard, K D; Heller, S R; Amiel, S A; Evans, M L; Dunger, D B; Hovorka, R.
Afiliação
  • Thabit H; Wellcome Trust-MRC Institute of Metabolic Science, University of Cambridge, Cambridge, UK.
Diabetes Obes Metab ; 17(5): 452-8, 2015 May.
Article em En | MEDLINE | ID: mdl-25492378
AIMS: To compare overnight closed-loop and sensor-augmented pump therapy in patients with type 1 diabetes by combining data collected during free-living unsupervised randomized crossover home studies. METHODS: A total of 40 participants with type 1 diabetes, of whom 24 were adults [mean ± standard deviation (s.d.) age 43 ± 12 years and glycated haemoglobin (HbA1c) 8.0 ± 0.9%] and 16 were adolescents (mean ± s.d. age 15.6 ± 3.6 years and HbA1c 8.1 ± 0.8%), underwent two periods of sensor-augmented pump therapy in the home setting, in combination with or without an overnight closed-loop insulin delivery system that uses a model predictive control algorithm to direct insulin delivery. The order of the two interventions was random; each period lasted 4 weeks in adults and 3 weeks in adolescents. The primary outcome was time during which sensor glucose readings were in the target range of 3.9-8.0 mmol/l. RESULTS: The proportion of time when sensor glucose was in the target range (3.9-8.0 mmol/l) overnight (between 24:00 and 08:00 hours) was 18.5% greater during closed-loop insulin delivery than during sensor-augmented therapy (p < 0.001). Closed-loop therapy significantly reduced mean overnight glucose levels by 0.9 mmol/l (p < 0.001), with no difference in glycaemic variability, as measured by the standard deviation of sensor glucose. Time spent above the target range was reduced (p = 0.001), as was time spent in hypoglycaemia (<3.9 mmol/l; p = 0.014) during closed-loop therapy. Lower mean overnight glucose levels during closed-loop therapy were brought about by increased overnight insulin delivery (p < 0.001) without changes to the total daily delivery (p = 0.84). CONCLUSION: Overnight closed-loop insulin therapy at home in adults and adolescents with type 1 diabetes is feasible, showing improvements in glucose control and reducing the risk of nocturnal hypoglycaemia.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Sistemas de Infusão de Insulina / Diabetes Mellitus Tipo 1 / Hipoglicemiantes / Insulina Tipo de estudo: Clinical_trials / Prognostic_studies Limite: Adolescent / Adult / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2015 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Sistemas de Infusão de Insulina / Diabetes Mellitus Tipo 1 / Hipoglicemiantes / Insulina Tipo de estudo: Clinical_trials / Prognostic_studies Limite: Adolescent / Adult / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2015 Tipo de documento: Article