Resection of calcified aortic heart leaflets in vitro by Q-switched 2 µm microsecond laser radiation.
Rohde, I; Masch, J-M; Theisen-Kunde, D; Marczynski-Bühlow, M; Bombien Quaden, R; Lutter, G; Brinkmann, R.
Afiliação
  • Rohde I; Institute of Biomedical Optics, University of Lübeck, Lübeck, Germany.
J Card Surg ; 30(2): 157-62, 2015 Feb.
Article em En | MEDLINE | ID: mdl-25530080
ABSTRACT

BACKGROUND:

Transcatheter aortic valve implantation (TAVI) can result in paravalvular leakage and stent deformation in the presence of severe calcification. This study was undertaken to determine the efficacy of laser-assisted resection of calcific aortic valve leaflets as a method to minimize the effects of calcium on perivalvular leakage during TAVI.

METHODS:

A Q-switched TmYAG laser emitting at a wavelength of 2.01 µm was used to evaluate the cutting efficiency on highly calcified human aortic leaflets in vitro (N = 10). A pulse energy of 4.3 mJ, a pulse duration of 0.8-1 µs, and a repetition rate of 1 kHz were used. The radiation was transmitted via a 200 µm core diameter quartz fiber. Resection was performed in a fiber-tissue contact mode on water-covered samples in a dish. The remnant particles were analyzed with respect to quantity and size by light microscopy.

RESULTS:

A resection rate of 40.4 ± 22.2 mg/min on highly calcified aortic leaflets was achieved. This corresponds to a cutting speed of approximately 1 cm/min; a laser dissection time of 3 min per leaflet is expected. The majority of the remnant particles (85.4%) were <6 µm in diameter, with only 0.1% exceeding 300 µm.

CONCLUSIONS:

The Q-switched TmYAG laser system showed promising results in cutting calcified aortic valves, by transmitting sufficient energy through a small flexible fiber. Catheter-based removal of aortic valve calcification may help to improve TAVI technology.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Valva Aórtica / Estenose da Valva Aórtica / Calcinose / Implante de Prótese de Valva Cardíaca / Terapia a Laser Limite: Humans Idioma: En Ano de publicação: 2015 Tipo de documento: Article