Your browser doesn't support javascript.
loading
Reliability of magnetic resonance elastography using multislice two-dimensional spin-echo echo-planar imaging (SE-EPI) and three-dimensional inversion reconstruction for assessing renal stiffness.
Low, Gavin; Owen, Nicola E; Joubert, Ilse; Patterson, Andrew J; Graves, Martin J; Glaser, Kevin J; Alexander, Graeme J M; Lomas, David J.
Afiliação
  • Low G; Department of Radiology, Addenbrooke's Hospital, University of Cambridge, England, United Kingdom.
  • Owen NE; University of Alberta, Edmonton, Alberta, Canada.
  • Joubert I; Division of Gastroenterology & Hepatology, Cambridge University Hospitals NHS Foundation Trust Hospital, England, United Kingdom.
  • Patterson AJ; Department of Radiology, Addenbrooke's Hospital, University of Cambridge, England, United Kingdom.
  • Graves MJ; Department of Radiology, Addenbrooke's Hospital, University of Cambridge, England, United Kingdom.
  • Glaser KJ; Department of Radiology, Addenbrooke's Hospital, University of Cambridge, England, United Kingdom.
  • Alexander GJ; Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA.
  • Lomas DJ; Division of Gastroenterology & Hepatology, Cambridge University Hospitals NHS Foundation Trust Hospital, England, United Kingdom.
J Magn Reson Imaging ; 42(3): 844-50, 2015 Sep.
Article em En | MEDLINE | ID: mdl-25537823
ABSTRACT

BACKGROUND:

To evaluate the reliability of MRE using a spin-echo echo-planar imaging (SE-EPI) renal MRE technique in healthy volunteers.

METHODS:

Institutional review board approved prospective study in which all participants provided written informed consent. Sixteen healthy volunteers comprising seven males and nine females with a median age of 35 years (age range 23 to 59 years) were included. Coronal 90 Hz and 60 Hz MRE acquisitions were performed twice within a 30-min interval between examinations. Renal MRE reliability was assessed by (i) test-retest repeatability, and (ii) inter-rater agreement between two independent readers. The MRE-measured averaged renal stiffness values were evaluated using intraclass correlation coefficient (ICC), Bland-Altman and the within-subject coefficient of variation (COV).

RESULTS:

For test-retest repeatability, Bland-Altman showed a mean stiffness difference between examinations of 0.07 kPa (95% limits of agreement -1.41, 1.54) at 90 Hz and 0.01 kPa (95% limits of agreement -0.51, 0.53) at 60 Hz. Coefficient of repeatability was 1.47 kPa and 0.52 kPa at 90 Hz and 60 Hz, respectively. The within-subject COV was 13.6% and 7.7% at 90 Hz and 60 Hz, respectively. ICC values were 0.922 and 0.907 for test-retest repeatability and 0.998 and 0.989 for inter-rater agreement, respectively (P < 0.001).

CONCLUSION:

SE-EPI renal MRE is a reliable technique.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Imageamento por Ressonância Magnética / Imagem Ecoplanar / Técnicas de Imagem por Elasticidade / Rim Tipo de estudo: Diagnostic_studies / Observational_studies Limite: Adult / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2015 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Imageamento por Ressonância Magnética / Imagem Ecoplanar / Técnicas de Imagem por Elasticidade / Rim Tipo de estudo: Diagnostic_studies / Observational_studies Limite: Adult / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2015 Tipo de documento: Article