Your browser doesn't support javascript.
loading
Time-resolved Characterization of Particle Associated Polycyclic Aromatic Hydrocarbons using a newly-developed Sequential Spot Sampler with Automated Extraction and Analysis.
Eiguren Fernandez, Arantzazu; Lewis, Gregory S; Spielman, Steven R; Hering, Susanne V.
Afiliação
  • Eiguren Fernandez A; Aerosol Dynamics Inc., 935 Grayson St., Berkeley, CA, phone: +1 (510) 649-9360, fax: +1 (510) 649-9361.
  • Lewis GS; Aerosol Dynamics Inc., 935 Grayson St., Berkeley, CA, phone: +1 (510) 649-9360, fax: +1 (510) 649-9361.
  • Spielman SR; Aerosol Dynamics Inc., 935 Grayson St., Berkeley, CA, phone: +1 (510) 649-9360, fax: +1 (510) 649-9361.
  • Hering SV; Aerosol Dynamics Inc., 935 Grayson St., Berkeley, CA, phone: +1 (510) 649-9360, fax: +1 (510) 649-9361.
Atmos Environ (1994) ; 96: 125-134, 2014 Oct 01.
Article em En | MEDLINE | ID: mdl-25574151
ABSTRACT
A versatile and compact sampling system, the Sequential Spot Sampler (S3) has been developed for pre-concentrated, time-resolved, dry collection of fine and ultrafine particles. Using a temperature-moderated laminar flow water condensation method, ambient particles as small as 6 nm are deposited within a dry, 1-mm diameter spot. Sequential samples are collected on a multiwell plate. Chemical analyses are laboratory-based, but automated. The sample preparation, extraction and chemical analysis steps are all handled through a commercially-available, needle-based autosampler coupled to a liquid chromatography system. This automation is enabled by the small deposition area of the collection. The entire sample is extracted into 50-100µl volume of solvent, providing quantifiable samples with small collected air volumes. A pair of S3 units was deployed in Stockton (CA) from November 2011 to February 2012. PM2.5 samples were collected every 12 hrs, and analyzed for polycyclic aromatic hydrocarbons (PAHs). In parallel, conventional filter samples were collected for 48 hrs and used to assess the new system's performance. An automated sample preparation and extraction was developed for samples collected using the S3. Collocated data from the two sequential spot samplers were highly correlated for all measured compounds, with a regression slope of 1.1 and r2=0.9 for all measured concentrations. S3/filter ratios for the mean concentration of each individual PAH vary between 0.82 and 1.33, with the larger variability observed for the semivolatile components. Ratio for total PAH concentrations was 1.08. Total PAH concentrations showed similar temporal trend as ambient PM2.5 concentrations. Source apportionment analysis estimated a significant contribution of biomass burning to ambient PAH concentrations during winter.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Risk_factors_studies Idioma: En Ano de publicação: 2014 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Risk_factors_studies Idioma: En Ano de publicação: 2014 Tipo de documento: Article