Your browser doesn't support javascript.
loading
CT pulmonary angiogram with 60% dose reduction: Influence of iterative reconstructions on image quality.
Pontana, F; Moureau, D; Schmidt, B; Duhamel, A; Faivre, J-B; Yasunaga, K; Remy, J; Remy-Jardin, M.
Afiliação
  • Pontana F; Service d'Imagerie Thoracique, Hôpital Calmette (EA 2694), CHRU et Université de Lille 2 Nord de France, boulevard Jules-Leclercq, 59000 Lille, France.
  • Moureau D; Service d'Imagerie Thoracique, Hôpital Calmette (EA 2694), CHRU et Université de Lille 2 Nord de France, boulevard Jules-Leclercq, 59000 Lille, France.
  • Schmidt B; Département Recherche et Développement; Siemens Healthcare, Forchheim, Germany.
  • Duhamel A; Service de biostatistiques (EA 2694); CHRU et Université de Lille 2 Nord de France, 59000 Lille, France.
  • Faivre JB; Service d'Imagerie Thoracique, Hôpital Calmette (EA 2694), CHRU et Université de Lille 2 Nord de France, boulevard Jules-Leclercq, 59000 Lille, France.
  • Yasunaga K; Service d'Imagerie Thoracique, Hôpital Calmette (EA 2694), CHRU et Université de Lille 2 Nord de France, boulevard Jules-Leclercq, 59000 Lille, France.
  • Remy J; Service d'Imagerie Thoracique, Hôpital Calmette (EA 2694), CHRU et Université de Lille 2 Nord de France, boulevard Jules-Leclercq, 59000 Lille, France.
  • Remy-Jardin M; Service d'Imagerie Thoracique, Hôpital Calmette (EA 2694), CHRU et Université de Lille 2 Nord de France, boulevard Jules-Leclercq, 59000 Lille, France. Electronic address: martine.remy@chru-lille.fr.
Diagn Interv Imaging ; 96(5): 487-93, 2015 May.
Article em En | MEDLINE | ID: mdl-25686775
ABSTRACT
GOALS To compare the quality of low-dose CT images with sinogram affirmed iterative reconstruction (SAFIRE), and full-dose CT with filtered back projection reconstructions (FBP). MATERIALS AND

METHODS:

Fifty pulmonary CT performed by a dual-source technique (120kVp; 110mAs) with (a) the same energy in both tubes, and (b) the distribution of reference mAs with 40% in tube A (44mAs) and 60% in tube B (66mAs). Each acquisition allowed reconstruction of (a) full-dose images (with both tubes) with FBP reconstructions (group 1); and (b) low-dose images (from tube A) reconstructed with SAFIRE (group 2).

RESULTS:

Group 2 images presented (a) a significant objective reduction in noise measured in the trachea on mediastinal (16.04±5.66 vs 17.66±5.84) (P=0.0284) and pulmonary (29.77±6.79 vs 37.96±9.03) (P<0.0001) images; (b) a similar subjective perception of noise and overall image quality (P=1), which was considered to be excellent in 66% (33/50) of the cases, with no influence on the detection of elementary pulmonary lesions of infiltration (98.4%; 95% CI=[96.9%-99.9%]).

CONCLUSION:

Despite a 60% reduction in radiation dose, the image quality with iterative reconstruction is objectively better and subjectively similar to full-dose FBP images.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Artéria Pulmonar / Doses de Radiação / Processamento de Imagem Assistida por Computador / Tomografia Computadorizada por Raios X Limite: Adolescent / Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2015 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Artéria Pulmonar / Doses de Radiação / Processamento de Imagem Assistida por Computador / Tomografia Computadorizada por Raios X Limite: Adolescent / Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2015 Tipo de documento: Article