Dosimetric feasibility of an "off-target isocenter" technique for cranial intensity-modulated radiosurgery.
Calvo-Ortega, Juan Francisco; Moragues, Sandra; Pozo, Miquel; Delgado, David; Casals, Joan.
Afiliação
  • Calvo-Ortega JF; Departamento de Oncología Radioterápica, Hospital Quirón, Barcelona, Spain. Electronic address: jfcdrr@yahoo.es.
  • Moragues S; Departamento de Oncología Radioterápica, Hospital Quirón, Barcelona, Spain.
  • Pozo M; Departamento de Oncología Radioterápica, Hospital Quirón, Barcelona, Spain.
  • Delgado D; Departamento de Oncología Radioterápica, Hospital Quirón, Barcelona, Spain.
  • Casals J; Departamento de Oncología Radioterápica, Hospital Quirón, Barcelona, Spain.
Med Dosim ; 40(4): 279-84, 2015.
Article em En | MEDLINE | ID: mdl-25824421
ABSTRACT
To evaluate the dosimetric effect of placing the isocenter away from the planning target volume (PTV) on intensity-modulated radiosurgery (IMRS) plans to treat brain lesions. A total of 15 patients who received cranial IMRS at our institution were randomly selected. Each patient was treated with an IMRS plan designed with the isocenter located at the target center (plan A). A second off-target isocenter plan (plan B) was generated for each case. In all the plans,100% of the prescription dose covered 99% of the target volume. The plans A and B were compared for the target dosage (conformity index [CI] and homogeneity index) and organs-at-risk (OAR) dose sparing. Peripheral dose falloff was compared by using the metrics volume of normal brain receiving more than 12-Gy dose (V12) and CI at the level of the 50% of the prescription dose (CI 50%). The values found for each metric (plan B vs plan A) were (mean ± standard deviation [SD]) as follows-CI 1.28 ± 0.15 vs 1.28 ± 0.15, p = 0.978; homogeneity index (HI) 1.29 ± 0.14 vs 1.34 ± 0.17, p = 0.079; maximum dose to the brainstem 2.95 ± 2.11 vs 2.89 ± 1.88Gy, p = 0.813; maximum dose to the optical pathway 2.65 ± 4.18 vs 2.44 ± 4.03Gy, p = 0.195; and maximum dose to the eye lens 0.33 ± 0.73 vs 0.33 ± 0.53Gy, p = 0.970. The values of the peripheral dose falloff were (plan B vs plan A) as follows-V12 5.98 ± 4.95 vs 6.06 ± 4.92cm(3), p = 0.622, and CI 50% 6.08 ± 2.77 vs 6.28 ± 3.01, p = 0.119. The off-target isocenter solution resulted in dosimetrically comparable plans as the center-target isocenter technique, by avoiding the risk of gantry-couch collision during the cone beam computed tomography (CBCT) acquisition.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Planejamento da Radioterapia Assistida por Computador / Encéfalo / Radiocirurgia Limite: Humans Idioma: En Ano de publicação: 2015 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Planejamento da Radioterapia Assistida por Computador / Encéfalo / Radiocirurgia Limite: Humans Idioma: En Ano de publicação: 2015 Tipo de documento: Article