Development and validation of an RP-HPLC method for the quantitation of odanacatib in rat and human plasma and its application to a pharmacokinetic study.
Police, Anitha; Gurav, Sandip; Dhiman, Vinay; Zainuddin, Mohd; Bhamidipati, Ravi Kanth; Rajagopal, Sriram; Mullangi, Ramesh.
Afiliação
  • Police A; Drug Metabolism and Pharmacokinetics, Jubilant Biosys Ltd, Industrial Suburb, Yeshwanthpur, Bangalore, 560 022, India.
  • Gurav S; Drug Metabolism and Pharmacokinetics, Jubilant Biosys Ltd, Industrial Suburb, Yeshwanthpur, Bangalore, 560 022, India.
  • Dhiman V; Drug Metabolism and Pharmacokinetics, Jubilant Biosys Ltd, Industrial Suburb, Yeshwanthpur, Bangalore, 560 022, India.
  • Zainuddin M; Drug Metabolism and Pharmacokinetics, Jubilant Biosys Ltd, Industrial Suburb, Yeshwanthpur, Bangalore, 560 022, India.
  • Bhamidipati RK; Drug Metabolism and Pharmacokinetics, Jubilant Biosys Ltd, Industrial Suburb, Yeshwanthpur, Bangalore, 560 022, India.
  • Rajagopal S; Drug Metabolism and Pharmacokinetics, Jubilant Biosys Ltd, Industrial Suburb, Yeshwanthpur, Bangalore, 560 022, India.
  • Mullangi R; Drug Metabolism and Pharmacokinetics, Jubilant Biosys Ltd, Industrial Suburb, Yeshwanthpur, Bangalore, 560 022, India.
Biomed Chromatogr ; 29(11): 1664-9, 2015 Nov.
Article em En | MEDLINE | ID: mdl-25914381
ABSTRACT
A simple, specific, sensitive and reproducible high-performance liquid chromatography (HPLC) assay method has been developed and validated for the estimation of odanacatib in rat and human plasma. The bioanalytical procedure involves extraction of odanacatib and itraconazole (internal standard, IS) from a 200 µL plasma aliquot with simple liquid-liquid extraction process. Chromatographic separation was achieved on a Symmetry Shield RP18 using an isocratic mobile phase at a flow rate of 0.7 mL/min. The UV detection wave length was 268 nm. Odanacatib and IS eluted at 5.5 and 8.6 min, respectively with a total run time of 10 min. Method validation was performed as per US Food and Drug Administration guidelines and the results met the acceptance criteria. The calibration curve was linear over a concentration range of 50.9-2037 ng/mL (r(2) = 0.994). The intra- and inter-day precisions were in the range of 2.06-5.11 and 5.84-13.1%, respectively, in rat plasma and 2.38-7.90 and 6.39-10.2%, respectively, in human plasma. The validated HPLC method was successfully applied to a pharmacokinetic study in rats.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Compostos de Bifenilo / Cromatografia Líquida de Alta Pressão / Cromatografia de Fase Reversa Tipo de estudo: Prognostic_studies Limite: Animals / Humans Idioma: En Ano de publicação: 2015 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Compostos de Bifenilo / Cromatografia Líquida de Alta Pressão / Cromatografia de Fase Reversa Tipo de estudo: Prognostic_studies Limite: Animals / Humans Idioma: En Ano de publicação: 2015 Tipo de documento: Article