Your browser doesn't support javascript.
loading
Accurate 3D temperature dosimetry during hyperthermia therapy by combining invasive measurements and patient-specific simulations.
Verhaart, René F; Verduijn, Gerda M; Fortunati, Valerio; Rijnen, Zef; van Walsum, Theo; Veenland, Jifke F; Paulides, Margarethus M.
Afiliação
  • Verhaart RF; a Hyperthermia Unit, Department of Radiation Oncology, Cancer Institute , Erasmus Medical Centre , Rotterdam , and.
  • Verduijn GM; a Hyperthermia Unit, Department of Radiation Oncology, Cancer Institute , Erasmus Medical Centre , Rotterdam , and.
  • Fortunati V; b Departments of Medical Informatics and Radiology , Biomedical Imaging Group Rotterdam, Erasmus Medical Centre , Rotterdam , the Netherlands.
  • Rijnen Z; a Hyperthermia Unit, Department of Radiation Oncology, Cancer Institute , Erasmus Medical Centre , Rotterdam , and.
  • van Walsum T; b Departments of Medical Informatics and Radiology , Biomedical Imaging Group Rotterdam, Erasmus Medical Centre , Rotterdam , the Netherlands.
  • Veenland JF; b Departments of Medical Informatics and Radiology , Biomedical Imaging Group Rotterdam, Erasmus Medical Centre , Rotterdam , the Netherlands.
  • Paulides MM; a Hyperthermia Unit, Department of Radiation Oncology, Cancer Institute , Erasmus Medical Centre , Rotterdam , and.
Int J Hyperthermia ; 31(6): 686-92, 2015.
Article em En | MEDLINE | ID: mdl-26134740
ABSTRACT

PURPOSE:

Dosimetry during deep local hyperthermia treatments in the head and neck currently relies on a limited number of invasively placed temperature sensors. The purpose of this study was to assess the feasibility of 3D dosimetry based on patient-specific temperature simulations and sensory feedback. MATERIALS AND

METHODS:

The study includes 10 patients with invasive thermometry applied in at least two treatments. Based on their invasive thermometry, we optimised patient-group thermal conductivity and perfusion values for muscle, fat and tumour using a 'leave-one-out' approach. Next, we compared the accuracy of the predicted temperature (ΔT) and the hyperthermia treatment quality (ΔT50) of the optimisations based on the patient-group properties to those based on patient-specific properties, which were optimised using previous treatment measurements. As a robustness check, and to enable comparisons with previous studies, we optimised the parameters not only for an applicator efficiency factor of 40%, but also for 100% efficiency.

RESULTS:

The accuracy of the predicted temperature (ΔT) improved significantly using patient-specific tissue properties, i.e. 1.0 °C (inter-quartile range (IQR) 0.8 °C) compared to 1.3 °C (IQR 0.7 °C) for patient-group averaged tissue properties for 100% applicator efficiency. A similar accuracy was found for optimisations using an applicator efficiency factor of 40%, indicating the robustness of the optimisation method. Moreover, in eight patients with repeated measurements in the target region, ΔT50 significantly improved, i.e. ΔT50 reduced from 0.9 °C (IQR 0.8 °C) to 0.4 °C (IQR 0.5 °C) using an applicator efficiency factor of 40%.

CONCLUSION:

This study shows that patient-specific temperature simulations combined with tissue property reconstruction from sensory data provides accurate minimally invasive 3D dosimetry during hyperthermia treatments T50 in sessions without invasive measurements can be predicted with a median accuracy of 0.4 °C.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Modelagem Computacional Específica para o Paciente / Neoplasias de Cabeça e Pescoço / Hipertermia Induzida Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Ano de publicação: 2015 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Modelagem Computacional Específica para o Paciente / Neoplasias de Cabeça e Pescoço / Hipertermia Induzida Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Ano de publicação: 2015 Tipo de documento: Article