Evaluation of a Spiral Groove Geometry for Improvement of Hemolysis Level in a Hydrodynamically Levitated Centrifugal Blood Pump.
Artif Organs
; 39(8): 710-4, 2015 Aug.
Article
em En
| MEDLINE
| ID: mdl-26146791
ABSTRACT
The purpose of this study is to evaluate a spiral groove geometry for a thrust bearing to improve the hemolysis level in a hydrodynamically levitated centrifugal blood pump. We compared three geometric models (i) the groove width is the same as the ridge width at any given polar coordinate (conventional model); (ii) the groove width contracts inward from 9.7 to 0.5 mm (contraction model); and (iii) the groove width expands inward from 0.5 to 4.2 mm (expansion model). To evaluate the hemolysis level, an impeller levitation performance test and in vitro hemolysis test were conducted using a mock circulation loop. In these tests, the driving conditions were set at a pressure head of 200 mm Hg and a flow rate of 4.0 L/min. As a result of the impeller levitation performance test, the bottom bearing gaps of the contraction and conventional models were 88 and 25 µm, respectively. The impeller of the expansion model touched the bottom housing. In the hemolysis test, the relative normalized index of hemolysis (NIH) ratios of the contraction model in comparison with BPX-80 and HPM-15 were 0.6 and 0.9, respectively. In contrast, the relative NIH ratios of the conventional model in comparison with BPX-80 and HPM-15 were 9.6 and 13.7, respectively. We confirmed that the contraction model achieved a large bearing gap and improved the hemolysis level in a hydrodynamically levitated centrifugal blood pump.
Palavras-chave
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Coração Auxiliar
/
Hemodinâmica
/
Hemólise
Tipo de estudo:
Evaluation_studies
/
Prognostic_studies
Limite:
Animals
Idioma:
En
Ano de publicação:
2015
Tipo de documento:
Article