ABSTRACT
INTRODUCTION:
The expanded use of implantable cardioverter-defibrillators (ICDs) to prevent sudden cardiac death (SCD) in hypertrophic cardiomyopathy (HCM) based on risk stratification in individuals without known previous ventricular arrhythmia is justified by an acceptable risk of device-related adverse events. Such complications, leading to surgical procedures or inappropriate electrical shocks, may impact mortality, morbidity, quality of life, and cost-effectiveness. METHODS ANDRESULTS:
From the Swedish ICD Registry, implants due to HCM since 1995 until November 2012 in patients aged ≥18 years were identified and medical records reviewed. Inappropriate ICD shock occurred in 14.3% (46 of 321 patients; mean follow-up 5.4 years) with a recurrent episode in 28.2% of them. In multivariable analysis, hazard ratio (HR) for atrial fibrillation was 3.5 (95% confidence interval 1.8-6.8; P < 0.001) but showed no significant association to male sex (HR = 0.77), age (HR = 0.99), secondary indication (HR = 1.02) or device, ICD-DR/CRTD vs. ICD-VR (HR 1.07). Inappropriate shocks were triggered by atrial fibrillation/flutter or ectopic tachycardia (56.5%), sinus tachycardia (14.5%), lead dysfunction (14.5%), and T-wave oversensing (13.0%). A reintervention, besides elective device replacement, occurred in 92 patients (totally 150 procedures). The majority were lead-related (70.0%) procedures, especially of the ICD lead. Reintervention was associated with female sex (HR = 1.6 P = 0.04).CONCLUSION:
Inappropriate ICD shock triggered by atrial arrhythmias, lead dysfunction, or complications requiring surgical interventions, is a concern in HCM patients who will be eligible for long-term prevention of sudden death. Efforts to avoid adverse events and provide balanced risk-benefit information are important, especially in primary prevention.Palavras-chave
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Cardiomiopatia Hipertrófica
/
Morte Súbita Cardíaca
/
Desfibriladores Implantáveis
/
Traumatismos por Eletricidade
Tipo de estudo:
Etiology_studies
/
Prevalence_studies
/
Prognostic_studies
/
Risk_factors_studies
Limite:
Female
/
Humans
/
Male
/
Middle aged
País/Região como assunto:
Europa
Idioma:
En
Ano de publicação:
2015
Tipo de documento:
Article