Your browser doesn't support javascript.
loading
Development and evaluation of analytical performance of a fully automated chemiluminescent immunoassay for protein induced by vitamin K absence or antagonist II.
Fujita, Kenju; Kinukawa, Hideki; Ohno, Kenichi; Ito, Yumiko; Saegusa, Haruhisa; Yoshimura, Toru.
Afiliação
  • Fujita K; Research and Development, Abbott Japan Co., Ltd., Chiba, Japan. Electronic address: Kenju.fujita@abbott.com.
  • Kinukawa H; Research and Development, Abbott Japan Co., Ltd., Chiba, Japan.
  • Ohno K; R&D Center for Diagnostic Reagents, Denka Seiken Co., Ltd., Niigata, Japan.
  • Ito Y; R&D Center for Diagnostic Reagents, Denka Seiken Co., Ltd., Niigata, Japan.
  • Saegusa H; R&D Center for Diagnostic Reagents, Denka Seiken Co., Ltd., Niigata, Japan.
  • Yoshimura T; Research and Development, Abbott Japan Co., Ltd., Chiba, Japan.
Clin Biochem ; 48(18): 1330-6, 2015 Dec.
Article em En | MEDLINE | ID: mdl-26210849
ABSTRACT

OBJECTIVES:

Protein induced by vitamin K absence or antagonist II (PIVKA-II), an abnormal form of prothrombin, has been used as an aid in the diagnosis of hepatocellular cancer (HCC) as a tumor marker. We developed a fully automated quantitative immunoassay for PIVKA-II on the ARCHITECT® i systems. The aim of this study was to evaluate the analytical performance of this assay. DESIGN AND

METHOD:

Assay imprecision, sensitivity, dilution linearity, high dose hook effect, sample type equivalency, assay interferences of potential interfering materials and correlation with Picolumi PIVKA-II (Eidia, Tokyo, Japan) were evaluated.

RESULTS:

The percentage coefficient of variation (%CV) of total imprecision ranged from 2.8% to 5.4% with 10 levels of samples. The limit of blank (LoB), limit of detection (LoD), and limit of quantitation (LoQ) were less than 0.63 mAU/mL, 1.62 mAU/mL, and 8.25 mAU/mL, respectively. Linearity up to 30,000 mAU/mL, no high dose hook effect, no difference among sample types and no interference of common drugs and endogenous substances were observed. Correlation study with the Picolumi PIVKA-II gave a correlation coefficient of 0.93 and a regression slope of 1.07.

CONCLUSIONS:

The results demonstrate that the fully automated prototype ARCHITECT PIVKA-II assay is an accurate, highly sensitive and precise assay for the measurement of PIVKA-II levels in human sera and plasmas.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Precursores de Proteínas / Protrombina / Imunoensaio / Biomarcadores Tumorais / Automação Laboratorial / Medições Luminescentes Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Ano de publicação: 2015 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Precursores de Proteínas / Protrombina / Imunoensaio / Biomarcadores Tumorais / Automação Laboratorial / Medições Luminescentes Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Ano de publicação: 2015 Tipo de documento: Article