[Flow cytometric test using eosin-5'-maleimide (EMA) labelling of red blood for diagnosis of hereditary spherocytosis].
Wang, Jiying; Zheng, Bin; Zhao, Yuping; Chen, Xuejing; Liu, Yan; Bo, Lijin; Zheng, Yizhou; Zhang, Fengkui; Ru, Kun; Wang, Huijun.
Afiliação
  • Wang J; Institute of Hematology and Blood Disease Hospital, CAMS & PUMC, Tianjin 300020, China.
  • Zheng B; Institute of Hematology and Blood Disease Hospital, CAMS & PUMC, Tianjin 300020, China.
  • Zhao Y; Institute of Hematology and Blood Disease Hospital, CAMS & PUMC, Tianjin 300020, China.
  • Chen X; Institute of Hematology and Blood Disease Hospital, CAMS & PUMC, Tianjin 300020, China.
  • Liu Y; Institute of Hematology and Blood Disease Hospital, CAMS & PUMC, Tianjin 300020, China.
  • Bo L; Institute of Hematology and Blood Disease Hospital, CAMS & PUMC, Tianjin 300020, China.
  • Zheng Y; Institute of Hematology and Blood Disease Hospital, CAMS & PUMC, Tianjin 300020, China.
  • Zhang F; Institute of Hematology and Blood Disease Hospital, CAMS & PUMC, Tianjin 300020, China.
  • Ru K; Institute of Hematology and Blood Disease Hospital, CAMS & PUMC, Tianjin 300020, China.
  • Wang H; Institute of Hematology and Blood Disease Hospital, CAMS & PUMC, Tianjin 300020, China.
Zhonghua Xue Ye Xue Za Zhi ; 36(7): 598-601, 2015 Jul.
Article em Zh | MEDLINE | ID: mdl-26304086
ABSTRACT

OBJECTIVE:

To investigate the sensitivity and specificity of eosin-5'-maleimide (EMA)assay for the diagnosis of hereditary spherocytosis (HS), and to verify the stability of reagent and samples.

METHODS:

EMA flow cytometry test, NaCl-osmotic fragility test and acidified glycerol lysis test were performed using peripheral blood samples from 80 patients with HS and 44 patients with other blood diseases, the sensitivity and specificity of the three methods were compared, and the feasibility of EMA binding test was estimated. The stability of EMA reagent and HS samples stored at different temperatures were tested.

RESULTS:

Among the 124 tested samples, the sensitivity and specificity of EMA binding test was 0.925 and 0.954, that of NaCl-osmotic fragility test was 0.950 and 0.455, and that of acidified glycerol lysis test was 1.000 and 0.318, respectively. Although the sensitivity of NaCl-osmotic fragility test and acidified glycerol lysis test was a little higher than that of EMA binding test, the specificity of the former two methods was poor, they couldn't clearly distinguish whether spherocytosis is hereditary spherocytosis. The experiment results showed that EMA was sensitive to the temperature and should not be stored in a small aliquots at -80 ℃ over a period of 6 months. The stability of the HS sample was better, 6 days storage at 4 ℃ and 3 days storage at room temperature had no influence on the results.

CONCLUSION:

EMA binding test by flow cytometry showed good sensitivity and specificity for HS diagnosis. EMA reagent should be stored at-80 ℃ and the HS samples should be tested within 6 days storage at 4 ℃ and 3 days at room temperature.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Esferocitose Hereditária / Anquirinas / Amarelo de Eosina-(YS) / Citometria de Fluxo / Testes Hematológicos Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: Zh Ano de publicação: 2015 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Esferocitose Hereditária / Anquirinas / Amarelo de Eosina-(YS) / Citometria de Fluxo / Testes Hematológicos Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: Zh Ano de publicação: 2015 Tipo de documento: Article