Metrological approaches to organic chemical purity: primary reference materials for vitamin D metabolites.
Nelson, Michael A; Bedner, Mary; Lang, Brian E; Toman, Blaza; Lippa, Katrice A.
Afiliação
  • Nelson MA; Material Measurement Laboratory, Chemical Sciences Division, National Institute of Standards and Technology, Gaithersburg, MD, 20899-8392, USA. michael.nelson@nist.gov.
  • Bedner M; Material Measurement Laboratory, Chemical Sciences Division, National Institute of Standards and Technology, Gaithersburg, MD, 20899-8392, USA.
  • Lang BE; Material Measurement Laboratory, Biosystems and Biomaterials Division, National Institute of Standards and Technology, Gaithersburg, MD, 20899-8392, USA.
  • Toman B; Information Technology Laboratory, Statistical Engineering Division, National Institute of Standards and Technology, Gaithersburg, MD, 20899-8392, USA.
  • Lippa KA; Material Measurement Laboratory, Chemical Sciences Division, National Institute of Standards and Technology, Gaithersburg, MD, 20899-8392, USA.
Anal Bioanal Chem ; 407(28): 8557-69, 2015 Nov.
Article em En | MEDLINE | ID: mdl-26345446
ABSTRACT
Given the critical role of pure, organic compound primary reference standards used to characterize and certify chemical Certified Reference Materials (CRMs), it is essential that associated mass purity assessments be fit-for-purpose, represented by an appropriate uncertainty interval, and metrologically sound. The mass fraction purities (% g/g) of 25-hydroxyvitamin D (25(OH)D) reference standards used to produce and certify values for clinical vitamin D metabolite CRMs were investigated by multiple orthogonal quantitative measurement techniques. Quantitative (1)H-nuclear magnetic resonance spectroscopy (qNMR) was performed to establish traceability of these materials to the International System of Units (SI) and to directly assess the principal analyte species. The 25(OH)D standards contained volatile and water impurities, as well as structurally-related impurities that are difficult to observe by chromatographic methods or to distinguish from the principal 25(OH)D species by one-dimensional NMR. These impurities have the potential to introduce significant biases to purity investigations in which a limited number of measurands are quantified. Combining complementary information from multiple analytical methods, using both direct and indirect measurement techniques, enabled mitigation of these biases. Purities of 25(OH)D reference standards and associated uncertainties were determined using frequentist and Bayesian statistical models to combine data acquired via qNMR, liquid chromatography with UV absorbance and atmospheric pressure-chemical ionization mass spectrometric detection (LC-UV, LC-ACPI-MS), thermogravimetric analysis (TGA), and Karl Fischer (KF) titration.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Espectrometria de Massas / Vitamina D / Espectroscopia de Ressonância Magnética / Cromatografia Líquida Tipo de estudo: Guideline / Prognostic_studies / Risk_factors_studies Idioma: En Ano de publicação: 2015 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Espectrometria de Massas / Vitamina D / Espectroscopia de Ressonância Magnética / Cromatografia Líquida Tipo de estudo: Guideline / Prognostic_studies / Risk_factors_studies Idioma: En Ano de publicação: 2015 Tipo de documento: Article