Effect of delayed scanning on imaging and on the diagnostic accuracy of vertical root fractures in two photostimulable phosphor plates digital systems.
Nascimento, H A R; Visconti, M A P G; Ferreira, L M; Suarez, M A; Haiter Neto, F; Freitas, D Q.
Afiliação
  • Nascimento HA; Division of Oral Radiology, Department of Oral Diagnosis, Piracicaba Dental School, State University of Campinas, São Paulo, Brazil. arnhelena@gmail.com.
  • Visconti MA; Division of Oral Radiology, Department of Oral Diagnosis, Piracicaba Dental School, State University of Campinas, São Paulo, Brazil.
  • Ferreira LM; Division of Oral Radiology, Department of Oral Diagnosis, Piracicaba Dental School, State University of Campinas, São Paulo, Brazil.
  • Suarez MA; Division of Oral Radiology, Department of Oral Diagnosis, Piracicaba Dental School, State University of Campinas, São Paulo, Brazil.
  • Haiter Neto F; Division of Oral Radiology, Department of Oral Diagnosis, Piracicaba Dental School, State University of Campinas, São Paulo, Brazil.
  • Freitas DQ; Division of Oral Radiology, Department of Oral Diagnosis, Piracicaba Dental School, State University of Campinas, São Paulo, Brazil.
Int Endod J ; 49(10): 973-9, 2016 Oct.
Article em En | MEDLINE | ID: mdl-26354080
ABSTRACT

AIM:

To evaluate the influence of delayed scanning on images obtained with two PSPs digital systems and on the diagnostic accuracy of vertical root fracture (VRF) by means of objective and subjective analyses.

METHODOLOGY:

Forty single-rooted human teeth were divided into two groups, one without VRFs and another with VRFs induced by a universal testing machine. Two digital systems (VistaScan(®) and Express(®) ) were used to radiograph all teeth, and the resulting plates were scanned at four time-points T0-immediately, T1-30 min, T2-2 h and T3-4 h after exposure. An aluminium (Al) wedge was used to evaluate the change in mean grey values as each scan was delayed. Three observers screened all images for VRFs, and one-fourth of the sample was revaluated after thirty days. Areas under the receiver operating characteristic (ROC) curve, sensitivity, specificity and accuracy values were compared by anova.

RESULTS:

Intra- and interobserver agreement ranged from moderate to substantial and fair to moderate, respectively. There was no significant difference amongst scan delays with regard to sensitivity, specificity and accuracy; however, there were significant differences in the area under the ROC curve, with the 4-h delayed scan being associated with lower values compared to the others (P = 0.019). As for objective analysis, there was a significant difference amongst all different scanning time-points for the two systems (P = 0.001), except between the 30-min and 2-h delayed scans in the VistaScan(®) system.

CONCLUSION:

Whilst delayed scanning caused changes to the density of images acquired with the systems studied, it did not seem to interfere with VRF diagnosis except when scanning was delayed for 4 h, which should therefore be avoided.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Fraturas dos Dentes / Raiz Dentária / Radiografia Dentária Digital Tipo de estudo: Diagnostic_studies / Evaluation_studies / Prognostic_studies Limite: Humans Idioma: En Ano de publicação: 2016 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Fraturas dos Dentes / Raiz Dentária / Radiografia Dentária Digital Tipo de estudo: Diagnostic_studies / Evaluation_studies / Prognostic_studies Limite: Humans Idioma: En Ano de publicação: 2016 Tipo de documento: Article